ARCUS®digma Handheld.
A 5.1 Locating the headgear
Place headbow @ on the patient`s head.
Locate the rear supports £ over and
behind the ears.
Lock the rubber band
5
.
Adjust the nasal support ” so that the
receiver # does not collide with the
patient`s nose.
• Orientation of the headgear to any of the
reference points is not necessary.
• Make sure the headgear is securely seated.
• If necessary, secure the headbow addition
ally with a rubber band.
A 5.2 Attaching the upper-jaw bite-
fork
Position the upper-jaw bitefork | in the
patient`s mouth so that the numbers 1 and
2 on the upper surface \ are visible.
For taking impressions of the upper
dentition, silicone such as Registrato
Fa. Vaco have provided the best results.
@
”
£
#
\
|
5