6.2 Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Clean manually or in a thermodisinfector only!
6.2.1 Cleaning: Manual cleaning - external
Accessories required:
▪ Tap water 30 °C ± 5 °C (86 °F ± 10 °F)
▪ Brush, e.g. medium-hard toothbrush
▶
Brush off under flowing tap water.
Preparation methods according to ISO 17664
33