Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing
20 / 26
▶
Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
Note
Do not place the motor in disinfectant baths!
7.2.4 Manual internal disinfection
Not applicable.
7.2.5 Manual drying
NOTICE
Moisture.
Corrosion and other product damage.
▶
Store motors in a dry place.
▶
If there is any residual moisture after cleaning and disinfecting, soak up the
residual moisture with a soft cotton cloth and then wipe dry.
▶
Carefully apply low doses of compressed air to the inside and outside of the
disassembled sleeve until no droplets of water are visible any longer.
7.3 Automated reprocessing
Not applicable.
7.4 Care products and systems - Servicing
The bearing sites of the INTRA LUX motors are provided with a high-quality,
lifelong lubrication.
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
NOTICE
Excess care products leaking from the straight or contra-angle handpiece (oil,
cleanser and disinfectant) can penetrate the motor and cause it to malfunction.
▶
Please make sure to always follow the care instructions supplied with the
instruments.
▶
Remove the motor sleeve and wipe off the inside and outside once a week.
▶
Wipe off the outside of the open motor with a dry cloth.