Instructions for use

INTRAflex LUX 2 Programm 
2307 LN - 0.535.6190, 2320 LN - 0.535.6180,
2325 LN - 0.535.6210

Содержание 0.535.6180

Страница 1: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 ...

Страница 2: ...istributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Страница 3: ... LN or 2320 LN 16 5 4 Removing INTRA LUX head from shanks 2307 LN or 2320 LN 17 5 5 Insert the milling cutters or diamond grinders in contra angle 2325 L 17 5 6 Removing the cutting tools or diamond grinder from the contra angle 2325 LN 18 6 Troubleshooting 19 6 1 Replacing the O rings on the coupling on the supply hose 19 6 2 Cleaning the spray nozzle 19 7 Reprocessing steps in accordance with IS...

Страница 4: ...8 Tools and consumables 25 9 Terms and conditions of warranty 26 Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 Table of contents 4 28 ...

Страница 5: ...eduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAVO 888 275 5286 Email techservice kavokerr com www kavo com Target group This document is intended for dentists and dental office staff The startup sec tion is also intended for service technicians General mar...

Страница 6: ...onditions Temperature range Transportation and storage conditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage The warning notes are designated as shown below DANGER In cases which if not prevented directly le...

Страница 7: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 1 User instructions NOTICE In cases which if not prevented could lead to material damage 7 28 ...

Страница 8: ...e product and accessories appropriately before disposal If there is any injury to soft tissue do not continue treatment in the oral cavity with compressed air driven instruments 2 2 Technical condition A damaged device or components could injure patients users and third parties Only operate devices or components if they are undamaged on the outside Check that the device is working properly and is ...

Страница 9: ...touch soft tissue with the handpiece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of preparation After treatment place the medical device properly in the cradle without the tool 2 5 Service and repair Repairs servicing and safety checks may only be performed by trained service personnel The follow...

Страница 10: ...ations for use This medical device is intended for dental treatment only All other types of use or modifications of the product are not permitted and can be hazardous The medical device is intended for the following applications Removal of carious material cavity and crown preparations removal of fillings processing of tooth and restora tion surfaces A medical device according to relevant national...

Страница 11: ...elf the patient and third parties from hazards to prevent contamination from the product 3 2 Technical Specifications 3 2 1 Reducing shank 2307 LN Identification 1 green ring Transmission ratio 2 7 1 Drive pressure 2 5 to 3 0 bar 29 to 44 psi Air consumption 50 to 60 Nl min Speed range 5 000 to 12 000 min 1 Idle torque at 3 0 bar 44 psi 3 2 Ncm The values apply for the application of the INTRA hea...

Страница 12: ...to 3 0 bar 29 to 44 psi Air consumption 50 to 60 Nl min Speed range 60 000 to 140 000 min 1 Idle torque at 3 0 bar 44 psi 0 32 Ncm Handpiece cutters or grinders can be used Attachable to all MULTIflex LUX MULTIflex LED couplings 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to 70 C 4 F to 158 F Relative humidity 5 RH to 95 RH absence of conde...

Страница 13: ...ir Contaminated and moist cooling air can cause malfunctions Make sure that the supply of cooling air is dry clean and uncontaminated according to ISO 7494 2 4 1 Mount the MULTIflex LUX MULTIflex LED coupling WARNING Detachment of the medical device during treatment A medical device that is not properly locked can release from the MULTIflex coupling during treatment Before each use check if the me...

Страница 14: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 4 Start up and shut down 4 2 Checking the amount of water 14 28 ...

Страница 15: ...731 between the coupling and the medical device and check the following pressures ð Drive pressure drive pressure recommended return air pressure spray water pressure and spray air pressure Drive pressure 2 5 to 3 0 bar 29 to 44 psi Recommended drive pressure 3 0 bar 44 psi Return air pressure 1 bar 15 psi Spray water pressure 1 0 to 2 0 bar 15 to 29 psi Spray air pressure 1 5 to 2 5 bar 22 to 36 ...

Страница 16: ...LED of the coupling or reduce its service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it backward until the coupling audibly locks in the medical device Pull on the medical device to make sure that it is securely affixed to the coupling 5 2 Removing the medical devic...

Страница 17: ...ers with worn or damaged shafts Risk of injury tool may fall out during treatment Never use dental burs or diamond grinders with damaged or worn shafts CAUTION Danger of injury from cutters or grinders Infections or cuts Wear gloves or finger stalls CAUTION Hazard from defective chuck system The cutter or grinder could fall out and cause injury Pull on the cutter or grinder to check that the chuck...

Страница 18: ...le 2325 LN WARNING Hazard from rotating dental bur or diamond grinder Lacerations and damage to the chuck system Do not touch the dental bur or diamond grinder when it is rotating Do not press the push button while the dental bur or diamond grinder is ro tating Remove the dental bur diamond grinder from the contra angle handpiece after treatment to avoid injury and infection when putting it away N...

Страница 19: ...y be lubricated with a cotton ball wetted with KaVo Spray Press the O ring between your fingers to form a loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle WARNING Hazard from nonsterile products Infection danger to the care provider and patient Reprocess and sterilise the medical device properly before the next use CAUTION Hazard fro...

Страница 20: ... tool Reprocess the medical device as soon as possible after treatment To minimise the risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composite or blood immediately Do not place in solutions or similar substances 7 2 Manual reprocessing WARNING Sharp tool in the medical device Injury hazard from sharp and or p...

Страница 21: ...nds the following products based on compatibility of the materi als The microbiological efficacy must be ensured by the disinfectant manufac turer and proven by an expert opinion Approved disinfectants CaviWipes and CaviCide made by Metrex Consumables required Cloths for wiping the medical device Spray the disinfectant on a cloth then wipe down the medical device and allow the disinfectant to act ...

Страница 22: ... KaVo CLEANpac bag and place it on the corresponding care adapter Press the spray key for 1 to 2 seconds Care of clamping chuck Note Only possible for the INTRAflex contra angle KaVo recommends servicing the chucking system once weekly Remove the tool from the medical device Position the tip of the spray nipple in the opening and apply the spray Press the spray key for 1 to 2 seconds 7 4 2 Servici...

Страница 23: ...ogether with the guide bush of the chuck to be ser viced against the tip of the service coupling Press the button marked with the chuck service symbol Note Close the chuck service mode Option 1 Place the dental handpieces in the QUATTROcare PLUS 2124 A close the front door and start theservice procedure Option 2 After three minutes with no service procedure running the device automatically switche...

Страница 24: ...l lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minutes at 135 C 275 F Drying time 16 min Steriliser using the gravity method at least 10 minutes at 135 C 275 F Drying time 30 min Steriliser using the gravity method at least 60 minutes at 121 C 250 F Drying time 15 min Remove contra angle handpieces and turbine handpieces from the steam steriliser immediately after co...

Страница 25: ... shanks 2307 LN 2320 LN INTRAflex LUX contra angle handpiece 2325 LN Instrument stand 3 005 5204 X X Insert for IN TRAflex 3 005 3089 X X Cellulose pad 100 units 0 411 9862 X X Cleanpac 10 units 0 411 9691 X X Nozzle pin 0 410 0931 X X Spray repair set 2163 0 410 0610 X Material summary Mat no KaVo Spray USA and Canada 2113 A 0 411 9660 QUATTROcare plus Spray USA and Canada 2141 P 1 005 4524 Chuck...

Страница 26: ...not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manufactu...