Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390
6 Troubleshooting | 6.1 Checking for malfunctions before initial start-up
6 Troubleshooting
6.1 Checking for malfunctions before initial start-up
NOTICE
The coupling system is fitted with light.
Defects on the light source.
▶
Before startup, check the glass rod light conductor for damage.
▶
If the glass rod light conductor is damaged, do not use the coupling any
longer and have it repaired by the service staff.
NOTICE
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
Note
Stop working if the O-ring is missing or damaged.
6.2 Troubleshooting
6.2.1 Exchanging the O-rings
NOTICE
Improper care of the O-rings.
Malfunction or complete failure.
▶
Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wetted
with KaVo Spray.
▶
Replace the O-rings, if the coupling is leaky.
▶
Press the O-ring between your fingers to form a loop.
▶
Shove the O-ring to the front, and remove it.
▶
Insert new O-rings into the grooves and spray with KaVo Spray.