5 Operation | 5.1 Connecting the coupling to the hose

5 Operation

Note

At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.

5.1 Connecting the coupling to the hose

Plug the MULTIflex LUX Midwest coupling onto the turbine hose and secure
it tightly with the union nut screw of the hose.

Spray the O-rings lightly with KaVo Spray.

The coupling remains screwed onto the hose.

5.2 Removing the coupling from the hose

In order to remove the coupling rotate the sleeve of the hose in anticlock-
wise direction and take the coupling off the hose.

5.3 Attaching instruments

 WARNING

Detachment of the medical device during treatment.

A medical device that is not properly locked can detach from the coupling dur-
ing treatment.

Before each use, check if the medical device is securely locked onto the
coupling.

NOTICE

Inaccurate coupling can destroy the glass rod of the coupling or reduce its ser-
vice life.

Make sure that the handpiece is accurately coupled and firmly seated on
the coupling.

Accurately plug the instrument with Midwest connector onto the
MULTIflex LUX Midwest coupling and push it backward until the coupling
can be heard to snap into place in the medical device.

Check the secure fit of the instrument on the coupling by pulling on it.

11 / 20

Содержание 0.553.1390

Страница 1: ...Instructions for use MULTIflex LUX coupling Midwest 0 553 1390...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...shooting 13 6 1 Checking for malfunctions before initial start up 13 6 2 Troubleshooting 13 6 2 1 Exchanging the O rings 13 7 Reprocessing steps in accordance with ISO 17664 14 7 1 Preparations at the...

Страница 4: ...KaVo Technical Service If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavokerr com Target group This document is intended for...

Страница 5: ...dity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage The warning notes a...

Страница 6: ...osal 2 2 Technical condition A damaged device or components could injure patients users and third parties Only operate devices or components if they are undamaged on the outside Check that the device...

Страница 7: ...re authorised to do this Service technicians of KaVo branches after the appropriate product training Service technicians of KaVo authorised dealers after the appropriate product training Observe all t...

Страница 8: ...d by expert users and for the specified application exclusively You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these...

Страница 9: ...conditions NOTICE Startup after refrigerated storage Malfunction Prior to startup strongly refrigerated products must be allowed to warm up to a temperature of 20 C to 25 C 68 F to 77 F Temperature 2...

Страница 10: ...so 2 7 Reprocessing steps in accordance with ISO 17664 Page 14 WARNING Dispose of the product in appropriate manner Infection hazard Reprocess and sterilise the product and accessories before disposal...

Страница 11: ...e 5 2 Removing the coupling from the hose In order to remove the coupling rotate the sleeve of the hose in anticlock wise direction and take the coupling off the hose 5 3 Attaching instruments WARNING...

Страница 12: ...uctions for use MULTIflex LUX coupling Midwest 0 553 1390 5 Operation 5 4 Removing instruments 12 20 5 4 Removing instruments Hold the coupling tight and pull the instrument forward while twisting sli...

Страница 13: ...staff NOTICE Missing or damaged O rings Malfunction and premature failure Make sure that all O rings are on the coupling and are undamaged Note Stop working if the O ring is missing or damaged 6 2 Tro...

Страница 14: ...reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composite or blood immediately Do not place in solutions or similar substances 7 2 Manua...

Страница 15: ...e product and let it work ac cording to the disinfectant manufacturer Follow the instructions for use of the disinfectant 7 2 4 Manual disinfection internal Not applicable 7 2 5 Manual drying Blow off...

Страница 16: ...0 553 1390 7 Reprocessing steps in accordance with ISO 17664 7 4 Sterilisation 16 20 7 4 Sterilisation Non sterilisable 7 5 Storage Reprocessed products should be stored protected from dust with mini...

Страница 17: ...or use MULTIflex LUX coupling Midwest 0 553 1390 8 Tools 8 Tools Available from dental suppliers Material summary Mat No Key 0 411 1563 Spare washer 0 553 4762 O ring large 1 004 2776 O ring small 1 0...

Страница 18: ...nd their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corr...

Страница 19: ......

Страница 20: ...0 589 2002 bd 20190809 08 en...

Отзывы: