DISCHARGE POWER SUPPLY MANUAL

MODEL DC3005, MODEL 30010, MODEL 15012

selects between the output VOLTAGE, or CURRENT adjustment for the far 

right ecoder knob.  The left display shows the Discharge voltage setpoint or 

readback in volts. The right display shows the Discharge current setpoint or 

readback in amps. The Discharge voltage or current is adjustable using the 

adjust knob on the front panel. This knob will also adjust the setpoint when 

the unit is in Standby mode. 

Three indicator lights are located between the power switch and the Enable/

Standby button.  The top light indicates if the power limit is being exceeded.  

The middle light indicates remote operated interlock is open. And the bottom 

light indicates the unit is shutdown in remote mode.

2.3.2 Rear Panel

Figure 2-2 shows the back panel of the Discharge Power Supply. An RS-232 

connection is provided on the back panel of this Power Supply for remote 

operation.  RS-232 connection at the back panel utilizes a DB-9 connector.  

The RS-485 is for KRI autocontroller. 

A ground stud is provided on the rear panel of this unit for grounding the 

chassis of the Power Supply.

Analog connections, utilizing DB-25 female connectors, are provided on the 

back panel of this controller for remote operation.

2.4 Installation

2.4.1 Mounting

The Power Supply is designed to be mounted in a standard 19-inch (483 

mm) rack mount cabinet.  The height of this Power Supply is 1.75 inches (45 

mm) and the depth of the unit is approximately 21 inches (533 mm).  When 

installing the Discharge Power Supply into the cabinet refer to the Ion Source 

Controller Manual for installation orientation relative to the other ion source 

Power Supplies.  Install the Power Supply in the rack mount cabinet prior to 

making any electrical connections.  

2.4.2 Cooling

This Power Supply is forced air cooled with the air

fl

 ow entering and exiting 

through the back panel of the chassis.  Ensure that there are no obstructions 

in the rack mount cabinet that will restrict the air

fl

 ow.

Содержание DC3005

Страница 1: ...nson Inc 1330 Blue Spruce Drive Fort Collins Colorado 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com Copyright 2011 by Kaufman Robinson Inc All rights reserved No part of this pub...

Страница 2: ...TION AND INSTALLATION 2 1 2 2 Inventory 2 1 2 3 Physical Description 2 1 2 3 1 Front Panel 2 1 2 3 2 Rear Panel 2 2 2 4 Installation 2 2 2 4 1 Mounting 2 2 2 4 2 Cooling 2 2 2 4 3 Electrical connectio...

Страница 3: ...ww ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Figures Fig 2 1 Discharge Power Supply front panel 2 4 Fig 2 2 Discharge Power Supply rear panel 2 5 Tables Table 1...

Страница 4: ...485 High Speed Serial link for proprietary communications with KRI Controllers Fan on Demand cooling with thermal shutdown protection Front panel selectivity for local remote mode 1 2 Safety Caution O...

Страница 5: ...e for the Discharge Power Supply is 85 275 volts ac single phase at 50 60 Hz The maximum input current to the unit is electronically limited to 28 amps RMS Operation at lower line voltages may be limi...

Страница 6: ...30 Blue Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Table 1 1 Maximum Power Volt...

Страница 7: ...ny damage has occurred contact Kaufman Robinson Inc in addition to the ship ping company to report any damage see Warranty section 5 Retain packaging materials for shipment of the Power Supply Invento...

Страница 8: ...igure 2 2 shows the back panel of the Discharge Power Supply An RS 232 connection is provided on the back panel of this Power Supply for remote operation RS 232 connection at the back panel utilizes a...

Страница 9: ...s purchased with an Ion Source and Fila ment Power Supply interconnecting cables are provided Information regard ing installation of these connecting cables can be found in the Ion Source Filament Cat...

Страница 10: ..._____________ Copyright 2011 by Kaufman Robinson Inc 1330 Blue Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC30...

Страница 11: ...Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 163 7 21 1 167 7 21 Input Power RS...

Страница 12: ...pply will be in Standby with the Dis charge Amps and Volts Setpoints displayed The Setpoints that are displayed will be the last setpoints used before the supply was switched OFF If the last operation...

Страница 13: ...lts Discharge Amps display Displays the discharge current setpoint and run val ues in amps When the yellow setpoint indicator below the display is illuminated the setpoint value is displayed When the...

Страница 14: ...of methods Some type of terminal or communications program or programming can be used for controlling this unit Some examples of programs that can be used are Hyper Terminal provided with Microsoft Wi...

Страница 15: ...the unit is functioning normally or a number that corresponds to the front panel help codes as described in diagnostics section 4 MEASure CURRent or MEAS CURR Measure current query Returns the DC out...

Страница 16: ...e return NOTE If either VX command is zero the unit will go to standby 3 2 4 Operational sequence Using a terminal program configured for the communications settings previ ously mentioned input the fo...

Страница 17: ...t VA0 then press the enter key The out put will drop to zero and the Standby LED will be illuminated Output Current Monitoring of the output current can be done using the following command Input RD0 t...

Страница 18: ...Supply for connection to the analog interface The pin descriptions for the interface are given below in table 1 There is a button on the front panel that can be used to switch from remote to local co...

Страница 19: ...on nected to common The shutdown is deactivated when 5 volts is applied or no contact is made Remote shutdown is indicated on the front panel when activated 15 Auxiliary 5 volts DC 200 mA max Also ava...

Страница 20: ...ruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Table 3 4 Interlock connections Pin Nu...

Страница 21: ...gas bottle to the ion source Power must be removed from the Power Supplies prior to performing maintenance on the Ion Source 4 2 Diagnostic Table The following table may be used to assist in determin...

Страница 22: ...e PC Shows up only in re mote RS 232 operation Display says HLP 8 for 5 seconds Informational display only No action taken on unit operation Unit recovers from error condi tion if the problem goes awa...

Страница 23: ...ts KRI s obligation under this Limited Warranty is at KRI s option to repair or replace any defective equipment or parts of the equipment without charge to the purchas er which are returned shipping p...

Страница 24: ...oblem encountered or the item to be returned Product Serial number Detailed description of problem Date of purchase Name and address of company Contact person If return to KRI is required you will be...

Отзывы: