KATflow 210
5 Commissioning
5.6 Heat quantity measurement (HQM)
Where equipped, heat quantity (energy) and heat flow (energy flow) can be
measured.
If a heat quantity unit is specified for the Process Value, the KF210 will prompt the
user for the Specific Heat Capacity of the medium in J/g/K (for example 4.186 J/g/K
for water).
The Output options menu for the PT100 will allow the user to select the
temperature input source; either PT100 temperature sensors or a fixed value for
measurement against a known inlet or outlet temperature. Where PT100 sensors
are selected, the Wizard will prompt the user for a temperature offset, which may
be useful where the temperature of the medium differs from the temperature of the
pipe wall (for example with unlagged pipes). If a fixed value is selected, the user will
be asked to specify this value.
When heat quantity units are selected, these behave as any other Process Value
and may be totalized, logged, or applied to a Process Output.
5.7 Sound velocity measurement (SVM)
The measured sound velocity (SOS) is available as a diagnostic function during
measurement and may be applied to a Process Output by selecting “C” from the
appropriate output menu.
5.8 Dual-channel flow calculations (maths functions)
Where suitably equipped, dual channel calculations are available from the
System/Calculation/Math menu.
These allow the user to select the sum, difference, average (mean) or maximum of
the two flow channels.
This value may be displayed or applied to a Process Output by selecting MATH
from the appropriate output menu.
5.9 Wall thickness measurement (WTM)
Optional sensor probes to measure pipe wall thickness are available. The KF210
will recognise a connected probe when entering the Setup or WTG Wizards, the
measurement mode or the Scope function. Use the Setup Wizard or Installation
menu to set the pipe material. Select “Start Measurement”. The KF210 will recog-
nise the probe and display the measurement screen. Wall thickness will be shown
when the sensor is in good acoustic contact with the pipe.
5.9.1 Wall Thickness Gauge (WTG) Wizard
To confirm pipe thickness and sound speed, select the "WTG Wizard" from the
"Quick Start" menu. Enter the approximate expected thickness as "Reference THK"
and select "Calibrate".
37