KATflow 210
4 Operation
Select transducer configuration (number of
passes) using cursor keys.
Auto
Automatically recommended
1
1 pass, diagonal mode
2
2 passes, reflection mode
3
3 passes, diagonal mode
4
4 passes, reflection mode
..etc.
Confirm by pressing <
ENTER
>.
Use cursor keys to select
Start Measurement
.
Confirm by pressing <
ENTER
>.
If sensors are recognised, the serial number
will be shown. If not recognised or not connec-
ted, they may be selected from a list.
Sensor placement screen: Mount transducers
with suggested spacing and use middle bar for
fine adjustment of position (central position is
desired). Observe signal-to-noise (upper bar)
and quality (lower bar). These should be of
identical length.
Confirm by pressing <
ENTER
> to obtain
measurements.
Success!
4.5 Measurements
4.5.1 Main process value (PV) display
Measurement is started using the Quick Setup Wizard. Once all the parameters are
programmed, any subsequent power-on sequences will bring up the main PV dis-
play immediately.
Display screen
Operation
The main process value can be changed
using the menu structure.
Press
<ESC>
at any time to return to the
main menu.
Change to the totalizer display, the 3-line
display, and dual display mode by pressing
<NEXT>,
or to the diagnostic screens by
pressing
<DISP>.
24
Table 4: Quick setup wizard