3. Measurement Mode
16
3.2 How to Collect Data at Certain Time Intervals [Interval Mode]
Display
Key
Procedure
MENU
(1)
Press
MENU
key.
△
,
▽
SET
(2)
Select
2. MEASUREMENT
.
Then press
SET
key.
△
,
▽
SET
(3)
Select
2. INTERVAL
.
Then press
SET
key.
△
,
▽
SET
SET
(4)
Set the measurement condition
in the same manner as when
configuring Average Mode
measurement (P14).
When the measurement condition is configured, select
7. SET
OK!
, and press
SET
key.
☆
To select the item to be
changed
△
,
▽⇨
SET
☆
To change setting value
and YES/NO.
△
,
▽⇨
SET
☆
Setting is complete.
7. SET OK!
⇨
SET
(a)
Data at each point is average value.
POINT 1
···
AVE 1=
Σ
DATA(N)/N
POINT 2
···
AVE 1=
Σ
DATA(N)/N
·
·
·
·
·
·
POINT N --- AVE H=
Σ
DATA(N)/N
AVE=(AVE1+AVE2+ -----AVE N)/No. of POINT (N)
MAX=AVE I, MIN=AVE J
AVE1, AVE2,
···
AVE N
are stored as a data at each point.
Calculation Result
< MENU >
1. MONITOR
2
. MEASUREMENT
3. DATA OUTPUT
4. MEMORY CLEAR
5. UTILITY
< MEASUREMENT >
1. AVERAGE
2
. INTERVAL
3. FLOW RATE
< INT >
1. S-TIME(S) 01
2. DATA (N) 010
3. INT (min) 060
4. POINTS 024
5. MEMORY YES
6. PRINT YES
7
. SET OK!
How long it takes to finish measuring
INT (min) × No. of POINTS (N)
Time length required to measure at one point
S-TIME (S) × N (No. of Data)
DATA (1) (2) (3)
·······
(N)
DATA (1) (2) (3)
·······
(N)
DATA (1) (2) (3)
·······
(N)
Time interval until the next sampling starts
Interval Time --- INT (min)
[POINT 1]
[POINT 2]
[POINT N]
S-TIME (S)
<Measurement Setting
Screen>