Number  

name  

Function  

 

 

 

 

inlet  

Sampling air suction port. A clean pipe can also be attached to 

the suction port for measurement.  

Never block the suction opening during measurement, as this 

will result in malfunction.  
Please do so.  

 

 

 

Analog  

output  
terminal  

DC power input, analog output terminal, and alarm output 

terminal. See "2.4 Analog Output Terminal  
Connection" for detailed functions.  

 

 

 

 

RS485 

port (e.g. 

LAN port)  

MODBUS RTU communication connection terminal; connect a USB-

RS  485  converter  (converter  +  cable);  the  PC  side  can  set 

parameters and read counts of this unit. (Refer to  
"2.5 RS 485 Terminal Connection" for details.  

 

 

 

 

power jack  

DC power jack. Use the supplied AC adapter. See "2.1.1 Adapter 

Use" for details.  

 

 

 

 

LAN connector  

MODBUS RTU communication connection terminal. Connect a router 

or POE switch to this terminal, which allows the PC side to set 

parameters and read counts of the unit.  
See "1.4 LAN Connection" for details.  

 

 

 

PARTICLE. 

lamp  

LED is orange; LED flashes each time a particle is detected. The 

light turns off when no pa-particles are  
detected.  

 

 

STATUS 

lamp  

When the flow rate and LD status are normal, the LED is lit green. If 

the flow rate or LD is abnormal, the lamp will light up red.  

 

 

 

exhaust port  

3720-06 is treated by a built-in filter to discharge the gas.  
Never block the exhaust port during measurement, as this  
will result in malfunction.  

Содержание 3720-06

Страница 1: ...Laser Particle Sensor MODEL 3720 06 Intense Light Particle Sensor User s manual...

Страница 2: ...Zero filter 1 Manual 1 optional Extras ITEM model volume USB RS485 converter 3720 06 20 1 RS232C RS485 converter 3720 06 21 1 Constant velocity suction probe 3905 07 1 Pump Unit 3720 06 22 IP65 Prote...

Страница 3: ...er that requires caution The subject of each caution is illustrated inside the triangle e g the symbol shown to the left is the high temperature caution Indicates a prohibition Do not take the prohibi...

Страница 4: ...ny liquid enters the instrument turn the power off immediately remove the battery and disconnect the power cable It may result in electric shock fire or damage to the instrument Contact your distribut...

Страница 5: ...1 1 Adapter used 5 2 1 2 DC power supply used 5 2 1 3 POE Switch 6 2 2 Parameter Setting 6 2 3 Check condition of vacuum source 7 2 4 Analog output terminal connection 7 2 5 RS485 pin connection 9 2...

Страница 6: ...ta can be output with a current of 4 20 mA Can be connected to the EP CRMS system for long term measurements in a clean room environment Status lamps can indicate LD errors and flow errors Alarm outpu...

Страница 7: ...RS 485 Terminal Connection for details power jack DC power jack Use the supplied AC adapter See 2 1 1 Adapter Use for details LAN connector MODBUS RTU communication connection terminal Connect a rout...

Страница 8: ...please refer to the following instructions Connection method The following is the connection method between this product the USB RS 485 converter and a PC How to connect RS485 terminal andUSB RS485 co...

Страница 9: ...m Connection method The connection between the product and the router is shown below Shows the connection between this product and the POE switch ROUTER PC POE Switch Router PC If power is to be suppl...

Страница 10: ...50 60Hz Output DC 24V 2 1A Plug specifications Internaldiameter2 1mm 2 1 2 Use DC power supply Pin 7 of the analog pins can be used for Power 15 28V and Pin 8 can be used for DC power supply by using...

Страница 11: ...rs Sampling time setting 1 second step set in the range of 1 60 seconds All data within the sampling time period is totaled and output via analog output 4 20 mA This sampling time is common to all cha...

Страница 12: ...bps 8 bit none and 2 bit Error output setting When flow rate or LD error is detected current output is performed or not performed ON The current output is changed according to the contents of the abn...

Страница 13: ...The contact specification for this terminal is 60 V 0 4 A or less Use within this range 3 ALM ch3 Channel 3 measurement particle count or concentration output upon alarm detection terminal The contact...

Страница 14: ...of particles or concentration of channel 3 measurement Output range is 4 20mA 2 This output terminal is also used when a main unit flow error or LD error is detected The current output range for a fl...

Страница 15: ...tput current value of the analog output terminal alarm output state etc If the problem is detected investigate the cause of the problem by referring to 5 If the problem persists contact the place of p...

Страница 16: ...00 000CNT ft3 Data Storage 5000 pcs Setting Parameter Sampling time 1 sec 60 sec Output range setting set for each particle size Count 0 10CNT 0 100CNT 0 1 000CNT 0 10 000CNT Concentrations 0 10CNT cf...

Страница 17: ...95 RH or less no condensation External dimensions of main unit 180 W x 110 D x 190 H mm Weight 3 4 kg Standard accessories Manual 1 volume AC adapter 1 pc zero filter 1 set optional goods USB RS485 c...

Страница 18: ...source piping Or clogged pipe Examine the vacuum source for leaks Eliminate clogged areas be just about to Piping inside the unit is clogged Returns and Repairs 13 Torn internal parts of the main uni...

Страница 19: ...MS wireless sensors are communicating and measuring numbers are not stable misconnection Check the connection method 4 Circuit failure of main unit Return for repair 13 EP CRMS system misconfiguration...

Страница 20: ...he product is malfunctioning please read 5 Still if you re not feeling well Please contact the distributor Kanomax Japan K K see last page or the distributor where you purchased the product Repairs un...

Страница 21: ...g tel 024 23846440 83951688 83951788 fax 024 23898417 No 9 Zhenggun Road Douyi Economy Development District Shenyang TEL 024 89730178 fax 024 89730177 USA19 US Hwy 206 Andover New Jersey 07821 TEL 1 9...