3. Display Description and operation Procedure
29
(1)
Data Update Interval: The output data is output every one second.
(2)
Load Impedance:
1 K
Ω
or higher
(3)
Output Cable:
Please purchase our optional output cable (3442-01).
Method to output measurement data (in case of analog output)
Explanation
Concentration value
(mg/m
3
or CPM) will be
converted to the output
voltage.
<Alarm Setting>
When the reading exceeds the set concentration value during a
measurement, the alarm will be turned ON. And when the reading
becomes lower than the set concentration value, the alarm will be OFF.
This alarm will be activated only during a measurement.
(Even if the alarm is ON during a measurement, it will be OFF once the
measurement finishes.)
When alarm mode setting is OFF, the alarm function is not activated.
Use [
▲
] [
▼
] keys to change the mode setting.
After setting the alarm mode, press [SET] key.
The RANGE setting value will be highlighted. Use [
▲
] [
▼
] keys to set
the value. After setting, press [SET] key. Then the RANGE setting value
will stop being highlighted.
0
1
2
3
4
sec
(
Measurement Time
)
Voltage
* Use the dedicated output cable (3442-01) sold separately.
mg/m3
CPM
OFF
[
▲
] key to CPM setting
↓
↑
[
▼
] key to mg/m3setting
[
▲
] key to OFF setting
↓
↑
[
▼
] key to CPM setting
10.000
9.999
[
▼
] key to decrease
↓
↑
[
▲
] key to increase