8. Calibrating CO/CO
2
Sensor
38
8.3 Calibration Procedure – SPAN Calibration
Display
Procedure
Connect a Span gas cylinder to the IAQ Monitor as described in the
previous page.
Press
key to bring up the MENU screen.
Use
key to select “7.CALIBRATION”, and press key.
Use key to select the sensor that you want to calibrate
(“2.CO” or “3.CO2”), and press key.
(CO2 is selected in this example.)
Perform a calibration by using span gas.
Press key.
Use key to enter a concentration value displayed on the
SPAN gas cylinder, and press key.
After setting the concentration, select “YES” by using key
and press key.
Enter the concentration value shown on SPAN gas cylinder label.
*Concentration Setting Range
CO: 20-550ppm, CO2: 800-5500ppm
Reading before calibration