background image

USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370     

Page 10 of 16 

8.3 TROUBLE 

SHOOTING 

 

If the warming is not starting, check as follows:

 

  Power cable connected (UK only - and plug fuse OK)? 

  Power in the wall socket? 

  Cabinet fuses OK? 

  Power Switch on (If power is OK it will show a green light) 

  For older Cabinets, is the door properly closed? The door starts the warming by pressing 

in the safety switch located on the upper left side of the front panel. Pressing in this 
switch with your finger should start the warming if above switch is OK? 

  Is the fan rotating? It shall start as soon as you press in the safety switch as mentioned 

above. If it is not rotating, making a strange noise or rotates slowly it must be changed. 

  Is the heating element getting too warm? The heating element has a self resetting over 

temperature protection (85

°

C) and two 1000W heating wire loops. Only one of the 1000W 

heating wires is being used. If it breaks, the second can be used instead. If the 
temperature protection cuts out the element frequently, there is not enough airflow 
through the element. 

  Check the fan and that the air intake under the heating compartment is not blocked. 

 

NOTE: All checking that involves opening the heating compartment must be done by a 
qualified technician. 
 

ACCESSORIES AND SPARE PARTS 

 

ACCESSORIES

 

 

 Part 

no 

Description 

  Quantity

 

 GE-41500 

Shelf 

  1 

 GE-41600 

Basket 

  1 

 

SPARE PARTS

 

 

Part no 

Pos. in drawing 

Description   

Quantity

 

 

700-0180 

S2 

Mains power switch 

 

 

700-0181 

S1 

Safety switch at door 

 

1

(

old cabinets) 

 700-0183 

M1 

Fan 

  1

 700-0184 

HE 

Heating 

Element 

 1 

 700-0457 

G1 

Temperature 

sensor 

 1 

 700-0456 

Tr 

Transformer 

  1 

 700-0464 

T1 

Temperature 

regulator 

 1

(earlier  

model) 

 

700-0659 T1 

Temperature 

regulator   

1

(later model) 

 

700-0660 

 Diode 

1N4005  1

(later model) 

 

700-0661 

 

Capacitor 100 µF 63V 

 

1

(later model) 

700-0185 T2 

Thermostat 

 

 

1

 700-0187 

L1 

Red 

lamp 

  1

 

700-0201 

 

Glass for door to cabinet 175 cm high  1 

 

700-0211 

 

Glass for door to cabinet 90 cm high 

 700-0202 

 Handle 

for 

door 

 1

 700-0203 

 Wall 

mounting 

bracket 

 2 

 700-0204 

 Door 

gasket 

  2 

 700-0205 

 Set 

of 

screws  1 

 700-0206 

 Foot 

for 

cabinet 

 1 

 

700-0453 

 

Front panel label WC 1002, adhesive  1 

 700-0208 

 Fuse 

holder 

  1 

 700-0209 

F1 

Fuse 

T6,3A 

  10 

 700-0460 

F2 

Fuse 

T50mA 

  10 

Содержание GE-1380

Страница 1: ...ns important information concerning safety and daily use as well as maintenance instructions and should be stored for future use This manual is valid for all cabinets that are delivered after 2001 Manufactured by KANMED AB January 2004 PART NO GE 1002 070 3 ...

Страница 2: ...DESCRIPTION 4 3 DESCRIPTION OF FUNCTION 4 4 TECHNICAL DATA 5 5 UNPACKING AND PACKING LIST 6 6 INSTALLATION 6 7 USING THE CABINET 6 8 MAINTENANCE AND TROUBLE SHOOTING 8 9 ACCESSORIES AND SPARE PARTS 10 10 WARRANTY 11 11 CIRCUIT DIAGRAM FOR CABINETS DELIVERED UNTIL 2003 12 12 CIRCUIT DIAGRAM FOR CABINETS DELIVERED FROM 2004 13 ...

Страница 3: ...skets at the same time The cabinet can tilt forward if it is not secured to the wall Take extra caution if the cabinet is mounted on wheels especially when the cabinet is moved around Note If the cabinet is not secured to a wall don t pull out more than one shelf basket at a time Do not overload the shelves basket maximum 8 kilos per shelf Max 2 pcs of gel pads 45x50x1 5 Max load of the basket is ...

Страница 4: ...s are available Mount the rails with the screws supplied using the prepared mounting holes and then push the shelf basket in place by holding the front higher than the rear 3 DESCRIPTION OF FUNCTION The warming cabinet and its contents are warmed by warm air heated by a 1000 Watt warming element The hot air is circulated by a fan and distributed evenly through the outlets at the rear of the cabine...

Страница 5: ...r measure Height 13cm Width 52cm Dept 50cm Maximum load 20kg TEMPERATURES Working temperature 35 C 42 C selectable in steps of 1 C Of circulating air Over temperature protection Visual alarm 45 C Accuracy 1 C Hysteresis T2 overtemp 4 C max WARMING CAPACITY Warming of 10 gel pads GE 455015 3 hours 45 x 50 x 1 5 cm from 22 C to 40 MODE OF OPERATION Designed for continuous use CE MARKING According to...

Страница 6: ...NOTE Always secure the cabinets to a wall by using the built in top brackets to avoid the risk of tipping forward at unfortunate circumstances 7 USING THE CABINET Open the door and switch on the Cabinet with the green power switch at the upper right side Check the set temperature and adjust if necessary according to 7 2 7 1 GENERAL ADVICE The Cabinet is designed for continuous use and shall be lef...

Страница 7: ...OWN SET Press set button to show the set selected temperature The set value is shown and the out lamp blinks as long as the set button is pressed While the set button is pressed the set value can be altered by pressing the UP DOWN button UP Press for temperature rise DOWN Press for temperature decrease OUT Flashes as long as the set button is pressed Lights permanently while heating takes place re...

Страница 8: ...scription below NOTE To adjust the temperature T1 you have to have the door open On older cabinets equipped with a door switch the door switch S1 must be pressed with for example some tape while you do the correction Don t forget to remove the tape when you are finished PROCEDURE Place the external control sensor as described under GENERAL INFORMATION above Start the warming and wait until the tem...

Страница 9: ...ON ABOUT TEMPERATURE MEASUREMENT AND CALIBRATION The temperature sensor T2 is a capillary thermostat In the KANMED cabinet its hysteretic is 4 C max When delivered T2 is set to 2 C above the maximum value of T1 42 C The activation of T2 is identified by a clear click sound as well as by the red over temperature lamp is lit The procedure described below is a simplified method to check and if necess...

Страница 10: ...ut the element frequently there is not enough airflow through the element Check the fan and that the air intake under the heating compartment is not blocked NOTE All checking that involves opening the heating compartment must be done by a qualified technician 9 ACCESSORIES AND SPARE PARTS ACCESSORIES Part no Description Quantity GE 41500 Shelf 1 GE 41600 Basket 1 SPARE PARTS Part no Pos in drawing...

Страница 11: ...t apply if the product has been modified adjusted or repaired other than by KANMED or by organisations authorised by KANMED or modified adjusted or repaired not in accordance with written instructions provided by KANMED or if the equipment has been subject to misuse negligence or accident These warranties are made on the condition that prompt notification of a defect is given to KANMED or its auth...

Страница 12: ...red until 2003 1 Connection socket 2 F1 Fuse 6 3 A 3 S1 Safety switch activated by the door 4 S2 Mains switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuse ...

Страница 13: ...for cabinets delivered from 2004 1 Connection socket 2 F1 Fuse 6 3A 4 S2 Main switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuses 50mA ...

Страница 14: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 14 of 16 ...

Страница 15: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 15 of 16 ...

Страница 16: ...ICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 16 of 16 KanMed AB Gårdsfogdevägen 18B S 168 66 BROMMA SWEDEN Tel 46 0 8 564 80 630 Fax 46 0 8 564 80 639 E Mail info KANMED se Home page www KANMED se DISTRIBUTED BY ...

Отзывы: