Страница 1: ...Kane International Ltd Exhaust Gas Analyzer KANE EGA4 5 Distributed in the USA by...

Страница 2: ...PROBE BATTERIES 11 REPLACING BATTERIES 11 TIME AND DATE 11 BATTERY TYPE 11 3 2 3 1 3 3 CHARGING MIMH BATTERIES 11 BATTERY DISPOSAL 11 3 4 3 5 4 GENERAL SAFETY 12 5 FIRST TIME USE 13 6 DAILY USE 13 1...

Страница 3: ...CE 29 19 PRINTOUTS 30 31 14 15 WATER TRAP PARTICLE FILTER AND WATER STOP FILTER 32 33 19 KANE LINK WIRELESS MEASUREMENT AND DATA TRANFER 34 35 16 10 2 TEMPERATURE 10 3 AIRFLOW RH TEMPERATURE 10 3 1 KA...

Страница 4: ...4 KANE EGA4 5 MANUAL 19 COLD WEATHER PRECAUTIONS 39 19 WELCOME TO KANE CARE RECYCLING 20 40 17 SPECIFICATIONS 36 37 EU DECLARATION 38 18...

Страница 5: ...easy fit accessory clip on rear above the battery compartment Your analyzer flow detector system automatically detects a blockage in the sampling system Your analyzer prints test results using an opti...

Страница 6: ...ices to your analyzer When connected they stay connected until you use KANE LINK to remove them When powered on they automatically replace or add measurements to your analyzer See page 35 to add manag...

Страница 7: ...ANALYZER FEATURES AND KEYPAD KANE EGA4 5 MANUAL 7 2...

Страница 8: ...F2 Short press to Store F2 HOME Return to home screen UP Short press to scroll up DOWN Short press to scroll down BACK CANCEL BACK CANCEL OK ENTER OK ENTER 8 KANE EGA4 5 MANUAL Short press to print a...

Страница 9: ...nections see section 2 Battery charger connection Home Back Delete Scroll down Flue gas inlet Print F1 Inlet temperature T2 Exhaust gas probe temperature T1 Exhaust Gas inlet Exhaust gas probe tempera...

Страница 10: ...10 KANE EGA4 5 MANUAL Thermocouple Wire EP1 engine exhaust probe Battery compartment Exhaust Gas outlet 2 3 BACK OF ANALYZER PROBE...

Страница 11: ...REPLACING BATTERIES Turn over your analyzer remove battery compartment cover Fit 6 x NiMH AA rechargeable batteries with correct battery polarity Replace battery compartment cover TIME AND DATE After...

Страница 12: ...output of sensors present NOTE This is different from a calibration where your analyser is exposed to known gas mixtures but allowed to settle to a steady figure with readings adjusted to the test gas...

Страница 13: ...set up tools After powering on your analyzer choose tasks to perform using MENU Most tests can be made with little user activity Your analyzer status bar on the bottom of the displays shows current ti...

Страница 14: ...lect print destination Press ENTER key display changes to show print progress LOG A TEST REPORT Press STORE key until display shows LOG STORED To print logged data Select LOG ON in REPORTS menu Press...

Страница 15: ...OMMENTS MEASUREMENT Select task to perform SETTINGS Change date time measurement parameters alarms etc REPORTS Configure logging parameters view stored data ON SCREEN TRENDS Configure display trend in...

Страница 16: ...y change if all logs in memory are cleared TEMPERATURE Select desired option via UP DOWN OK to confirm SIMULATED NO2 Set assumed of NO2 present in sample for NOx Calculation 5 default ALARMS Set Toxic...

Страница 17: ...to exit without change MENU ITEM OPTIONS COMMENTS VIEW COMB REPORTS View reports VIEW PRS TEMP REPORTS View reports VIEW AIRFLOW REPORTS View reports VIEW KANE DTHA2 REPORTS View reports DELETE REPORT...

Страница 18: ...PTIONS COMMENTS SETUP Set SAMPLING INTERVAL TREND A Parameter TREND B Parameter TREND C Parameter TREND D Parameter START TREND A Start STATY TREND B Start START TREND C Start START TREND D Start STAR...

Страница 19: ...to view each page View current configuration setup Use Press HOME to exit change KANE EGA4 5 MANUAL 19 8 5 STATUS...

Страница 20: ...IL Enter operator owner information PRINTER Select IR printer type MANAGE LINK DEVICES Add or remove KANE LINK devices ALARM LEVELS Set alarm trigger levels for each gas sensor Note alarms set in incr...

Страница 21: ...For more accurate measurements SERVICE Restricted area for authorized personnel only KANE EGA4 5 MANUAL 21 MENU ITEMS OPTIONS COMMENTS MANUAL AIR ZERO Manually trigger Air Zero purge 8 7 TOOLS 8 8...

Страница 22: ...ration count down AUTOMATIC CALIBRATION During automatic calibration analyzer samples fresh air to zero toxic sensors set oxygen sensor to 20 95 After power on your analyser displays identity software...

Страница 23: ...ate time before making measurement Begins measurement process Use to manage options and start tests Typical exhaust gas analysis screen KANE EGA4 5 MANUAL 23 10 MEASUREMENT MENU 10 1 EXHAUST GAS Selec...

Страница 24: ...perature measurements 10 3 Your analyzer tests airflow RH temperature with an optional KANE DTHA2 wireless anemometer Requires KANE DTHA2 connected via KANE LINK See page 35 10 3 1 DTHA2 SCREEN Typica...

Страница 25: ...igh side low side pressure temperature in vehicle temperature humidity airflow on one screen See page 20 page 30 to add devices Manage Link devices Use to edit each device Press HOME to exit without c...

Страница 26: ...ODE Select required measurement parameters EXHAUST GAS AIRFLOW DURATION Choose test duration from 1 to 24 hours INTERVAL Choose sampling interval from 3 to 60 seconds TOTAL SAMPLES Indicates number of...

Страница 27: ...Typical test running screen Select KANE LINK or KANE APP modes Select text size for task screens 10 5 1 10 7 10 6 WIRELESS MODE READING SIZE TEST RUNNING KANE EGA4 5 MANUAL 27...

Страница 28: ...avoid exhaust dilution with back flushing ambient air Position hose assembly away from hot surfaces SAFETY WARNING Use suitable protection when disconnecting probe which may be hot Do not exceed anal...

Страница 29: ...r water trap Keep analyzer in line particle filter clean dry DO THIS EVERY TIME YOU USE THE ANALYZER Remove probe from exhaust TAKE CARE HOT PROBE allow to cool naturally Allow analyzer to purge in fr...

Страница 30: ...30 KANE EGA4 5 MANUAL Exhaust Gas Test Air Conditioning Test 14 PRINTOUTS...

Страница 31: ...KANE EGA4 5 MANUAL 31 Humidity KANE DTHA2 Temperature...

Страница 32: ...stop gas water vapour dust entering your analyzer However some boilers engines produce much higher volumes of water vapour which can affect your analyzer Your analyzer also has a hydrophobic water sto...

Страница 33: ...Stop filter WSF2 Particle filter PF2 Water trap SM50675 5 Replace particle filter Water stop filter Particle filter 6 Replace water stop filter Ensure small o ring is fitted when replacing water stop...

Страница 34: ...ng buttons Each serial number must be 10 digits long If shorter enter 0 s to make up to 10 eg enter serial number below as 0002001228 KANE WPCP TEMPERATURE PIPE CLAMP PROBE To add a temperature probe...

Страница 35: ...TOR To add a KANE79 select KANE79 using buttons Enter serial number using buttons Each serial number must be 10 digits long Use numeric part of serial number to pair your KANE LINK analyzer KANE LINK...

Страница 36: ...20 Hydrocarbon NDIR 1ppm 5 of reading or 12ppm volume 0 5000ppm Over range 10 000ppm Calculations Efficiency High C 0 1 1 of reading 0 119 9 Excess Air 0 1 0 2 of reading 0 119 9 Lambda AFR Petrol LPG...

Страница 37: ...idity 15 to 90 RH non condensing Ambient Operating Range 5 C to 50 C 10 to 90 RH non condensing Power Supply battery charger Input 110Vac 220 Vac nominal Output 12 VDC off load Physical Characteristic...

Страница 38: ...2 2013 IEC62321 1 2013 IEC62321 3 1 2013 IEC62321 5 2013 IEC62321 4 2013 IEC62321 7 2 2017 IEC62321 7 1 2015 IEC62321 6 2015 Signed for on behalf of Kane International Ltd 01 July 2022 Paul Morrison E...

Страница 39: ...happens readings may display as sensors may be permanently damaged If your analyzer is affected by condensation or water ingress leave running in a warm place with pump ON sampling fresh air for up to...

Страница 40: ...KANE CARE applies to any KANE analyser Service Recertification registered booked in via www anseddiagnostics com pages product registration or use QR code below 20 ANSED Diagnostic Solutions LLC 1528...