![Kane 9206 Quintox Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/kane/9206-quintox/9206-quintox_user-manual_3379473026.webp)
26
GAS UNITS:
Select ppm or ppm(n) or mg/m3 or mg/m3(n).
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
TEMPERATURE:
Allows the selection of Fahrenheit or Celsius.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
PRESSURE:
Allows the selection of pressure units.
Select from: mbar, In H
2
O, mm H
2
O, hPa, psi,.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET PERCENTAGE REFERENCE O
2
:
Can be set between 0% (equivalent to OFF) and 10%.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET NOX CALCULATION:
Select from: SUM, NO
2
or NO
Set REFERENCE NOx to the percentage required or defined by local regulations.
Typically 5% NO
2
is added to an NO reading. The value can be user set.
SUM adds the readings from an NO sensor and an NO
2
sensor when fitted
NO calculates an NOx reading from the NO reading where NOx = NO x 1.1
NO
2
calculates an NOx reading from the NO
2
reading where NOx = NO
2
x 2.05
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET COMPENSATION:
Select from YES or NO
CONVERSION FACTORS:
Display the Propane Equivalency Factor PEF and the Methane Equivalency Factor for the
IR module (if fitted) and the pitot factor – change from 0.10 to 1.00 as determined by the
pitot tube being used.
Содержание 9206 Quintox
Страница 1: ...1 KANE9206 Quintox Flue Gas Analyser Stock No 19332 5 November 2014 Kane International Ltd ...
Страница 44: ...44 Fit a new ribbon guiding the paper roll between the exposed ribbon and cartridge body Refit printer cover ...
Страница 49: ...49 ...
Страница 59: ...59 ...
Страница 71: ...71 ...