3. USING THE FOUR FUNCTION BUTTONS:
Switching ON
the Analyser
Press
/
button to switch the unit ON. This must be
done in fresh air to ensure that the analyser auto calibrates its
sensors properly.
When switched on, the analyser beeps and briefly displays battery
%, fuel and pressure units. Its bottom line counts down from 90
until the sensors are ready to use. If the analyser will not auto
calibrate, its sensors need to be replaced or recalibrated by an
authorised repair centre.
If an inlet temperature probe (optional) is connected into the T2
socket during its countdown, the measured temperature from the
inlet probe will be used as the inlet temperature.
If an inlet temperature probe is not connected to the analyser
during countdown the measured temperature from the flue probe
will be used as the inlet temperature.
If neither probe is connected during countdown the analyser’s
internal ambient temperature will be used as the inlet temperature.
Switching OFF
the Analyser
Press
/
button to switch the analyser OFF. The
display counts down from 30 with the pump on to clear the sensors
with fresh air – If the probe is still connected, make sure analyser
and probe are in fresh air.
Press
/
if you want to stop the countdown and return
to making measurements.
Note: The analyser will not switch off unless the CO
reading is below 20ppm.