KANE905 Operators Manual
Page 39
13. PRODUCT SPECIFICATION
Parameter
Resolution
Accuracy
Range
Temp Measurement
Flue Temperature
Inlet Temperature
0.1
C/F
0.1
C/F
+2.0
C +0.3% reading
+1
C +0.3% reading
0-1200
C/32 -2200
F
with suitable probe
0-50
C/32-122
F
Gas Measurement
Oxygen
Carbon Monoxide
H
2
compensated
Carbon Monoxide,
Nitric Oxide
Low range
Nitric Oxide
(optional)
Nitrogen Dioxide
(optional)
Sulphur Dioxide
(optional)
Pressure
Carbon Dioxide
*2
Losses
*2
Efficiency
*2
Excess Air
*2
Temp (Nett)
*2
CO/CO
2
ratio
*2
Poison Index
*2
0.1%
1ppm
1ppm
1ppm
1ppm
1ppm
1ppm
0.1mbar
0.1%
0.1%
0.1%
0.1%
1.0
C/F
0.0001
0.01%
+0.2%
*1
+5ppm <100ppm
+20ppm <400ppm
*1
+5% >400ppm
+20ppm <400ppm
*1
+5% <5000ppm
+10% >5000ppm
+3ppm <20ppm
+5ppm<100ppm
+5ppm <100ppm
*1
+5% >100ppm
+3ppm<20ppm
+5ppm<100ppm
+5ppm<100ppm
+5%>100ppm
+5.0% full scale
+0.3% reading
+1.0% reading
+1.0% reading
+0.2%
+2
C +0.3% reading
+0.0001
+0.01
0-25%
0-4000ppm
0-100000ppm
0-100ppm
0-5000ppm
100ppm
0-5000ppm
150 mbar
0-99.9%
0-99.9%
0-99.9%
0-2885.0%
0-1200
C/32-2200
F
0-0.9999
0-99.99
Pre-programmed Fuels
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil, Propane,
Butane, Anthracite, Coke, Coal, Kinsale Gas.
Dimensions
Weight
Handset
Probe
1kg
220mm x 55mm x 120mm
L240mm x Dia8mm with 285mm long stainless steel shaft, type
K thermocouple and 1.5m long neoprene hose
Ambient Operating Range
-5
C to +50
C/10% to 90% RH non condensing
Power Supply
(battery charger)
Input: 110Vac/220 Vac nominal
Output: 12 Vac off load
Battery Life
>8 hours from full charge
*1
Using dry gases at STP