background image

USING THE MENU 

 

Press 

 then navigate through the menu by using following keys… 

 

  = Scroll up 

 

 

= Scroll 
down 

 

  = Enter 

 

MENU 

SUB-MENU  OPTIONS / COMMENTS 

 

 

 

STORE 

VIEW 

Use the 

/

 and 

  keys to view a previously logged test. 

Other tests can then be viewed using the 

/

 keys. 

Press the menu key to exit. 

 

LOG 

Press the 

  key to manually log a test including date and time. 

Storage capacity is 255 tests. 

 

AUTO STORE 

Select YES to automatically log tests at a preset interval. 
The log interval can be set at between 2 and 99 seconds. 
Logging starts immediately after the log interval is set. 
To stop AUTO STORE either switch off or select menu / auto store 
/ NO. 

 

DEL ALL 

Clears all the memory when YES is selected 

 

EXIT 

 

 

 

 

SETUP 

 

 F 

The internal temperature of the meter can be displayed in 

C or 

 

AUTO OFF 

Auto power off timer in minutes.  Set at 00 to disable the auto 
power off. 

 

SET TIME 

Uses “Military” time.  7am = 07:00, 7pm = 19:00 

 

SET DATE 

Select format, e.g. DD-MM-YY, then set the date 

 

HEADER 

Sets printout header.  2 lines of up to 20 characters per line 

 

LANGUAGE 

English, French, Italian, German and Spanish 

 

CONTRAST 

Sets the display contrast. 

 

EXIT 

 

 

 

 

PRESSURE 

SMOOTH 

OFF = standard response.  ON = slow response to damp out 
pulses 

 

RESOLVE 

HIGH or LOW.  Selects the number of decimal places 

 

PS UNITS 

mbar, mmH

2

O, Pa, kPa, PSI, mmHg, hPa, inH

2

 

REPORT 

(3500-1/UK and 
3500-2/UK only. 
Also see section 5) 

TEST.  Starts the timed Let-by, Stabilisation and Tightness tests 
VIEW.  View all logged Let-by and Tightness tests 
DEL ALL.  Deletes all Let-by and Tightness tests 
SETTINGS.  Sets the duration of Let-by and Tightness tests in 
minutes 

 

EXIT 

 

  

 

 

SERVICE 

CODE 

For use by an approved service organisation only.  Leave set at 
‘0000’ 

 

 

 

EXIT 

 

 

 

Содержание 18965-2

Страница 1: ...KANE3500 PRESSURE METER Stock No 18965 2 June 2014 Kane International Ltd...

Страница 2: ...be live data frozen data or stored data 255 sets of tests can be stored in the non volatile memory Two lines of 20 characters can be added to the header of printouts The meter is controlled using 8 b...

Страница 3: ...he button is also a shortcut to the let by stabilization tightness test Only applies to KANE3500 1 UK and KANE3500 2 UK BOTTOM LINE OPTIONS The buttons change the bottom line of the display The bottom...

Страница 4: ...ot sure what the maximum pressure might be Always ensure that the meter you are using is correctly rated for the pressure that you intend to measure Excessive pressure 5 times nominal range can perman...

Страница 5: ...can be displayed in C or F AUTO OFF Auto power off timer in minutes Set at 00 to disable the auto power off SET TIME Uses Military time 7am 07 00 7pm 19 00 SET DATE Select format e g DD MM YY then set...

Страница 6: ...he meter in the orientation that it is being used This eliminates gravity effects on the very sensitive pressure transducer It is also recommend that the meter is switched on for at least five minutes...

Страница 7: ...EXAMPLE PRINTOUTS The standard printout is The MAX and MIN readings are those observed since last switch on NOTE Printouts of stored readings will also include the TEST NO below the header...

Страница 8: ...S 20 00 The stabilisation test counts down in seconds Duration is set via menu PS live pressure reading during the stabilisation period 8 TIME 00 PS 20 00 End of the stabilisation period Either press...

Страница 9: ...Let By Test and Tightness Test Printout NOTE Printouts of stored readings will also include the LOG NO Below the header...

Страница 10: ...r should be re calibrated and serviced annually by a Kane approved service centre Local regulations may require more frequent re calibration Calibration is performed in firmware and there are no user...

Страница 11: ...inal Pressure Ranges mbar mmH2O Pa kPa 80 00 800 0 8000 8 000 PSI mmHg hPa inH20 1 000 60 00 80 00 30 00 Model KANE3500 2 Nominal Pressure Ranges mbar mmH2O Pa kPa 160 00 1600 0 16000 16 000 PSI mmHg...

Страница 12: ...000 40 00 PSI mmHg hPa inH20 5 000 300 0 400 0 150 0 Model KANE3500 15 Nominal Pressure Ranges mbar kPa PSI 1040 0 100 00 15 00 mmHg hPa inH20 780 0 1040 0 400 0 Model KANE3500 30 Nominal Pressure Ran...

Страница 13: ...000 Pa 1 Pa 3 of reading 80 hPa 0 01 hPa 3 of reading Model KANE3500 2 Range Resolution Accuracy 20 Pa 0 3 Pa 2 Pa 2000 Pa 0 3 Pa 5 Pa 160 hPa 0 01 hPa 3 of reading Model KANE3500 5 Range Resolution A...

Страница 14: ...076 hPA 0 1 hPa 10 hPa 2070 0 hPA 1 hPa 3 of reading Dimensions Weight 310 grams with battery Handset 160mm x 80mm x 40mm 180mm incl spigots Ambient 0 C to 45 C Operating 10 to 90 RH non Range condens...

Страница 15: ...appropriate to check the meter before use by Use the normal start up sequence in the location where the meter will be used Switch on all localized electrical equipment capable of causing interference...

Страница 16: ...Specification EC EMC KI KANE3500 details the specific test configuration performance and conditions of use Please Note Batteries used in this instrument should be disposed of in accordance with curre...

Страница 17: ...e is recorded when each test is stored printed NOTE When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be necessary...

Страница 18: ...71JG Sending your KANE3500 Once the meter has been securely packed then your package is ready for shipment back to Kane If you do not have an account with a courier company you can take your package t...

Страница 19: ...Kane International Ltd Gibfield Park Avenue Atherton Manchester M46 0SY Southern Service Centre Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Northern Servic...

Страница 20: ......

Страница 21: ...rdshire AL7 1JG Your Details Name Job Title Company Name Company Address 1 Address 2 Town City County Postcode Country Phone Number Fax Number Mobile Number Email Address Product Details Note Proof of...

Страница 22: ...ut Kane Magazine Advert Trade Counter Literature Personal Recommendation Internet Exhibition Other Which do you read most often Often Sometimes Hardly Ever Registered Gas Engineer Gas Installer P H P...

Страница 23: ......

Страница 24: ...Thank you for buying this pressure meter Before use please register on our website www kane co uk or complete detach and return the Product Registration page...

Отзывы: