MID designations
Rated operation conditions/measuring ranges:
Calculator
q
: 10°C…180°C
DΘ
: 3K…170K
Temperature sensor pair
q
: 10°C…150°C
DΘ
: 3K…140K
Flow sensor
q
: 15°C…130°C
Mechanical environment:
M1 (fixed installation with minimum vibration).
Electromagnetic environment:
E1 and E2 (Domestic, light industrial and industrial). Signal cables from the meter
must be separated by at least 25cm distance to other installations.
Maintenance and repair:
The heat supplier is allowed to change communication module, battery,
temperature sensor pair and flow sensor. Sensor pair and flow sensor are
separately verified and can, therefore, be separated from the calculator. All repairs
require a following re-verification on an accredited laboratory.
Climatic environment:
The installation shall be made in non-condensing environments and in closed
location (indoor). The ambient temperature must be within 5…55°C.
MULTICAL® 601, type 67-B/C/D is suitable for temperature sensors type Pt500
MULTICAL® 601, type 67-A is suitable for temperature sensors type Pt100
Battery for replacement: Kamstrup type 66-00-200-100
MULTICAL® 601, type 67-A/B/C can be connected to flow sensor type ULTRAFLOW®,
electronic pick-up unit or flow sensor with reed switch output.
MULTICAL® 601, type 67-D must be connected to a flow sensor with 24 V active
pulse output.
Irrespective of flow sensor type, ”pulse/litres” must be identical on flow sensor and
calculator.