background image

User Manual

 

OBM-LT

 

10

 

 

4.

 INSTRUCTIONS  

    FOR USE 

 

 
4.1 Basic setup 

 

Step 1:

 Remove rubber cap and fill the bath with appropriate fluid (up to 80%) 

Step 2:

 

Connect the power cord into the power supply 230V±10%; 50Hz

 

Step 3:

 Put the main switch in position 1 (on the back of the device) 

Step 4:

 Press 

 button and

 

in 

Program Start Menu 

select the desired program. After selection press 

             

 button to start cycle 

Step 5:

 Check the fluid level and add if necessary 

Step 6:

 Press 

 button to stop the cycle  

 

 

4.2 Control panel with OLED display and keyboard 

 

 

 

 

START button 

ESC button 

STOP button 

OLED display 

Increase value button 

Decrease value button 

Navigate right button 

Navigate left button 

Navigate down button 

Navigate up button 

ENTER button 

MENU button 

Содержание OBM-LT

Страница 1: ...User Manual Calibration Bath OBM LT...

Страница 2: ...fits loss of income loss of business opportunities loss of use and other related exposures caused by incorrect use of product Warranty is valid for 13 months from date of shipping and includes spare p...

Страница 3: ...rt 12 4 2 4 Menu and navigation 15 4 3 Program Setup 16 4 4 Instrument Setup 18 4 4 1 Display Parameters 18 4 4 2 Time and Date 19 4 4 3 System Parameters 19 4 4 4 Bath Calibration 20 4 4 5 PID parame...

Страница 4: ...sure about safe use of the product in your specific case do not hesitate to contact your dealer or us via info kambic com for further information or opinions In no event shall Kambic d o o be held li...

Страница 5: ...in the bath DANGER Temperature fluid expansion may cause overspill Adjust fluid level RISK OF FAILURE Do not use rubber cap at elevated temperatures DANGER OF BURNS Use safety gloves when operating wi...

Страница 6: ...at is why we are using high capacity microprocessor controllers and different levels of heating and cooling Control panel with display is used to set the individual temperatures All controllers are al...

Страница 7: ...emperature Stability from 0 005 C to 0 05 C Temperature homogeneity Better than 0 05 C Display resolution 0 001 C Set up resolution 0 01 C Control MPC PID Power supply 10 230 V 50 Hz Wattage 650 W Com...

Страница 8: ...ar view Control panel Bath working chamber Rubber cap Refrigeration system compartment air intake Main switch Power cord connection RS 232 PC port USB port Tank drain plug Perforated back Front housin...

Страница 9: ...User Manual OBM LT 9 3 2 Rear view Bath working chamber Removable inner tank wall Openings on inner tank wall...

Страница 10: ...device Step 4 Press button and in Program Start Menu select the desired program After selection press button to start cycle Step 5 Check the fluid level and add if necessary Step 6 Press button to sto...

Страница 11: ...running 4 2 2 Main screen not running After button is pressed you see this screen Setpoint temperature Heating power Stirrer speed in Bath temperature Current program number Temperature status Bath te...

Страница 12: ...g Press button to enter Program Start Menu With and buttons select desired program Select from 0 to 10 Program 0 is reserved for manual program and 1 to 10 is automatic program After selection press o...

Страница 13: ...is not yet in setpoint tolerances Stable temperature is stable and it is in setpoint tolerances Current program the current running program number is 0 0 manual program 1 10 for automatic program Setp...

Страница 14: ...temperature current setpoint temperature 25 00 C setpoint range from 40 00 C to 130 00 C Stirrer speed the set speed for stirrer is 100 set range from 30 to 100 Step dwell time remaining time of temp...

Страница 15: ...enter MAIN MENU The Main Menu includes three submenus Program Setup Instrument Setup and Service Use navigation buttons and to select and then press button to enter selected submenu Press button to go...

Страница 16: ...to edit parameter With and buttons change value of parameter To move cursor left and right use and button Press button to confirm or button cancel changes After edit press button to go back to MAIN M...

Страница 17: ...start set range from 1 to maximum 10 if we have 10 temperature steps Cycling from step number must always be bigger than Cycling to step number 10 Cycle to step set the step number to which the cycli...

Страница 18: ...meter With and buttons change value of parameter To move cursor left and right use and button Press button to confirm or button cancel changes Press button to go back one level 4 4 1 Display Parameter...

Страница 19: ...format hh mm ss Date format yyyy mm dd 4 4 3 System Parameters Setup system parameters To change parameter you must first enter password Default values Max SETP limit 130 C Heat power limit 100 Compre...

Страница 20: ...0 C 0 500 C At 50 C 49 600 C At 100 C 99 600 C At 130 C 129 600 C 4 4 5 PID parameters Setup bath PID parameters To change parameter you must first enter password Default values PID Pb 1 25 C PID Ti 1...

Страница 21: ...User Manual OBM LT 21 4 4 6 Change Password For password setup first enter old password The factory default password is 1000 Now you can enter now password Confirmation screen for password has changed...

Страница 22: ...User Manual OBM LT 22 4 5 Service When you enter Service screen you see information of bath model bath serial number firmware version operating hours and after service hours...

Страница 23: ...leaned regularly to prevent a buildup of oil or dust For cleaning use some soft cleaning detergents and soft cloth Periodically check the fluid level in the bath to ensure that the level has not dropp...

Страница 24: ...Draining the Bath All our calibration baths are designed and equipped with tank drain plug for draining of the bath fluid The location of the drain plug is at back of the bath After drain screw back...

Страница 25: ...st wall It must also be in a dry and well ventilated room Additional space or relocation of the apparatus may be necessary when carrying out any maintenance or servicing The user must provide electric...

Страница 26: ...ume expansion at higher temperatures Fill the bath with appropriate fluid To avoid unlikable leakage do not overfill the bath The operator should also make sure that the bath contains enough fluid at...

Страница 27: ...t with your other household waste Instead you should protect human health and the environment by handing over your waste equipment to a designated collection point for the recycling of waste electrica...

Страница 28: ...9 ABBRIVATIONS TEMP chamber temperature SETPT set temperature PWR heating power PID proportional integral derivative control loop LED light emitting diode RS 232 standard communication interface USB...

Страница 29: ...s on refrigerant system check temperature sensor Increased noise check stirring check fluid Not able to reach high temperature check the max power on heater limit setting Not able to reach low tempera...

Страница 30: ...User Manual OBM LT 30 11 ATTACHMENTS 11 1 Wiring diagram 11 2 Declarations 11 3 RS232 Communication protocol...

Отзывы: