To connect the power for the CHA™ sensor, first check both wires from the source for polarity and voltage, then

label appropriately. Shielded twisted pair wire is recommended for both power and signal.

Connect positive wire to 24/12 VDC (+). Fig. 3-13

Connect negative wire to 24/12 VDC (-). Fig. 3-13

WARNING: Connecting a power source to the 4-20 mA ports on the TB will damage the 4-20 mA output and 

resulting failure of the unit.

Check voltage and polarity at terminal block.

1.

2.

3.

4.

KAM CHA

3939 Ann Arbor Drive

Houston, Texas

77063, USA

+1 713 784 0000

www.KAM.com

485 +

485 -

Loop2  In

Loop2 R

TN

Loop1 R

TN

Loop1 IN

GND

Vin       ( – )

Vin       ( + )

LED3

LED1

LED2

POWER    IN

J5

1

2

3

4

J7

ASTM

SA

YBOL

T

PT/CO

HAZE

FIG. 3-14

Serial Port

4-20 mA

Connections

{

{

{

COMMUNICATION WIRING

Connect the wiring to the desired outputs and/or serial port. Fig. 3-14 

For 4-20 mA outputs, LOOP1 will be set to Haze and LOOP2 to specified color range unless otherwise noted. To 

change selected ranges see pg. 14 of this manual.  

The RS485 serial port can be used for X,Y,Z data or MODBUS (see pg. 17 of this manual).

When wiring is complete ensure that the lid is properly placed back in place and that all bolts are tightened.

1.

2.

3.

4.

I N S TA L L AT I O N   C O N T I N U E D

Содержание Simple Precision CHA

Страница 1: ...84 0000 FAX 1 713 784 0001 Email Sales Kam com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 USA www KAM com User Manual KAM CHA COLORIMETER HAZE ANALYZER PTB 04 ATEX 1027 Ex II 2 G Ex db...

Страница 2: ...3 4 Operations and maintenance 15 Calibration 15 Color scale ranges 15 Hyperterminal RS485 15 Cleaning and inspection 16 Readings verification 16 5 MODBUS Configuration 17 CAUTION When installing the...

Страница 3: ...ean there are no moving parts Using long lasting LED light sources ensures long term stable performance with limited maintenance and power requirements In addition locating the electronics within an e...

Страница 4: ...ch is amplified and sent to the computer The computer in turn utilizes a mathematical formula to determine the exact color and haze turbidity of the fluid in the pipeline Final color measurement from...

Страница 5: ...0 600 Hazardous area PTB04 ATEX 1027 Ex II 2 G Ex db IIB T6 Gb NEMA4 IP66 equivalent Accuracy Haze 5 Color 1 Repeatability Haze 1 Color 1 Reproducibility Haze 1 Color 1 Haze Range 0 100 Color Range Co...

Страница 6: ...40 9 65 MM 213 218 218 MM 216 222 226 MM 226 232 240 MM 240 240 245 TABLE 2 2 FLANGE SIZE AND CLASS SL TABLE 2 3 DIMENSIONS A B C D E F INCHES 1 25 3 5 7 1 4 6 47 1 MM 32 90 180 117 12 25 Shaft Lengt...

Страница 7: ...6 CHAMANUAL 0917 KAM CONTROLS INC S P E C I F I C AT I O N S C O N T I N U E D INCHES 11 8 11 7 25 4 7 FIG 2 4 FLOWTHROUGH CHA A B C D MM 299 279 184 119 TABLE 2 5 DIMENSIONS C D A B ANSI FLANGES...

Страница 8: ...lled on a vertical line If the unit must be intalled on a horizontal line it should be at 3 or 9 o clock position Fig 3 2 The KAM CHA probe must be oriented so that the flow indicator is parallel to t...

Страница 9: ...n or removal The Seal Housing of the CHA sensor allows the probe to be inserted and removed from the pipe under pressure and flow conditions The CHA sensor should be inserted so that the window of the...

Страница 10: ...Wall Thickness 2 Example for D1 16 pipe WT 1 4 Seal Thickness 1 8 MID 16 1 4 1 8 2 MID 18 3 8 4 Measure the calculated MID from the top of the seal clamp and place a mark with a permanent marker or ta...

Страница 11: ...in probe damage include damage to sapphire windows FIG 3 11 9 Flow Direction Re tighten the Socket Cap Screws with a torque of 750 800 lb in or 62 5 to 66 6 lb ft Tighten the Hex Nuts on the top of t...

Страница 12: ...n top Locking Collar then slowly and with caution loosen the Socket Cap Screws Fig 3 12 NOTE Once the Socket Cap Screws have been loosened the CHA shaft may push out from the line Slide the CHA sensor...

Страница 13: ...in operation For installation using conduit All conduit runs must have a sealing fitting connected within one conduit diameter or 2 50mm whichever is less For installations in the EU all cable entries...

Страница 14: ...Ann Arbor Drive Houston Texas 77063 USA 1 713 784 0000 www KAM com 485 485 Loop2 In Loop2 RTN Loop1 RTN Loop1 IN GND Vin Vin LED3 LED1 LED2 POWER IN J5 1 2 3 4 J7 ASTM SAYBOLT PT CO HAZE FIG 3 14 Seri...

Страница 15: ...t two scales total The CHA does not require recalibration after changing scales In most cases LOOP1 will be set for haze and LOOP2 to the specified color scale CAUTION Inserting more than one jumper f...

Страница 16: ...you want for example CHA 1 Ignore the choose icon option Hit OK The Connect To dialogue box will now appear On the last line where it says Connect using choose the name of the communication port conn...

Страница 17: ...ately 2 5 Connect power to the CHA per instructions on pg 12 of this manual Connect to the RS 485 Serial Port using a RS485 to USB converter or RS485 to Serial converter Place CHA probe in a clear ves...

Страница 18: ...Red LED signal from reference fiber 40110 16 bit integer IR LED signal from reference fiber 40111 32 bit Float Blue LED signal transmission percent 40113 32 bit Float Green LED signal transmission per...

Отзывы: