23
OIDMANUAL 1219
KAM CONTROLS, INC.
M A I N T E N A N C E C O N T I N U E D
FIG. 5-5
(output data OID models serial numbers OID-15-1062 or higher))
Column 1
– Command number (for information only)
Column 2
– Time Code (for information only)
Column 3
– Offset value in counts. The factory offset should be around
6553 ±500. To convert the value to volts divide the value shown by 13107.
Column 4
– Signal value from the MPPC photodiode. This value is also in counts. It has range
from 6553 ± 500 to 58982 ( 0.5Vto 4.5V) depending on the product that the probe is seeing.
Column 5
– High Voltage supply going to the MPPC Photo diode. The range is
from 46.5VDC to 58.0VDC.
Column 6
– MPPC Photodiode Temperature, controlled by a thermoelectric cooler
inside the photodiode. The factory temperature is between 14C and 16C.
Column 7
– Reference Voltage that controls the High Voltage supply.
This value should be between .720 to .880VDC
Column 8
– Value of the photodiode monitoring the LED when it is
OFF. This value is shown in counts and should be less than 100.
Column 9
– Value of the photodiode monitoring the LED when it
is ON. This value is shown in counts and should be between 400
& 600. If the LED is damaged and not turning on this value will be
the same as in column 8.
Column 10
– Signal in counts after any compensations such
as compensation for temperature.
Column 11
– OID output on a 0-100% range
Column 12
– OID output on a 4-20mA range
Column 13
– Electronics temperature
Column 14
- CRC Cyclic Redundancy
check for information only