Kaixin 3451 Скачать руководство пользователя страница 7

Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet)                                                    User’s Manual V1.07 

- 1 - 

Chapter 1 Technical Specifications 

1.1 Technical parameters 

1. Gray scale: 256 
2. Monitor: 3.5” LCD 
3. Adapter rating: 100-240V~, 1.2-0.6A, 50-60Hz (model: BJE01-40-001M) 
4. Output of Adapter: DC12.8V 3.0A 
5. Main device rating: DC12V 3.0A 
6. Main Unit Size: approx. 90 * 98 * 25

 

(L * M * H, mm) 

7. Weight of main unit: approx. 620g (including battery and probe) 
 

1.2 Primary Functions 

1. System preset function; 
2. Mode conversion function; 
3. Adjustment of 

near 

field, far field, total gain and dynamic range; 

4. Frequency conversion function; 
5. Frame correlation function; 
6. Compress curve adjustment function; 
7. Color display function; 
8. Screen brightness and contrast adjustment function;   
9. Depth range selection function; 
10. Image freeze/unfreeze function; 
11. Measurement of backfat thickness, loin thickness and lean percentage; 
12. Image and data storage function; 
13. Energy saving; 
14. Chinese-English switch function. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 3451

Страница 1: ...KAIXIN MSU3 Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd ...

Страница 2: ...responsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts and accessories consumables involved repairs are Kaixin company original or approved by Kaixin 3 Related electrical equipment complies with national standards and...

Страница 3: ...e problems damages or losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic process Kaixin shall not be liable for any problems arising out of the process 7 Be sure to back up important data to external storage media such as not...

Страница 4: ... a potential danger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Up Direct current Keep dry Power switch Fragile Power supply indication Stacking limit by number Battery charge indicator Temperature limits Storage and transport Adapter connection indication Humid...

Страница 5: ...llation and Check 5 5 1 System installation 6 5 2 Install Remove the battery 6 5 3 Connection to shutter release 7 5 4 Connection to power 7 5 5 Ultrasonic probe check before and after operation 7 5 6 Main unit check before and after operation 7 5 6 1 Inspection before start up 7 5 6 2 Inspection after start up 8 5 7 System reset 8 Chapter 6 Functional Operation 9 6 1 Startup and Shutdown 9 6 2 Sy...

Страница 6: ...reeze 13 6 7 Image Data storage and viewing 13 6 7 1 Image data storage 13 6 7 1 1 Save image in B 2B mode 13 6 7 1 2 Save data in A mode 14 6 7 2 View image or data 14 Chapter 7 Fat thickness and Lean percentage measurement 15 7 1 Measurement items 15 7 2 Backfat measurement 15 7 3 Measurement of Swine s lean percentage 16 7 4 Measurement considerations 16 Chapter 8 System Maintenance 17 8 1 Main...

Страница 7: ...nit approx 620g including battery and probe 1 2 Primary Functions 1 System preset function 2 Mode conversion function 3 Adjustment of near field far field total gain and dynamic range 4 Frequency conversion function 5 Frame correlation function 6 Compress curve adjustment function 7 Color display function 8 Screen brightness and contrast adjustment function 9 Depth range selection function 10 Imag...

Страница 8: ...ning ultrasonic diagnostic instruments Vet are composed of main unit and probe etc 2 2 Components name Figure MSU3 main unit sketch map 2 3 Function keys description SN Function keys Function 1 Image parameters menu System setting menu in frozen status 2 Selection and adjustment of parametes Mode selection Depth selection etc 3 Freeze Unfreeze 4 Power switch 5 Press the Freeze switch to freeze unf...

Страница 9: ...3 Chapter 3 System Configuration 3 1 Standard Configuration 1 Main unit 1 unit 2 3 5MHz mechanical sector probe 1 PC 3 Power adapter 1 PC 4 Internal battery 1 PC 5 Power cable 1 PC 6 OTG U disk 1PC 7 Velcro straps 1 Set 8 Portable bag 1 PC 9 Shutter release 1 PC 3 2 Optional parts 1 Plastic seal box ...

Страница 10: ...o may result in explosion Warning Avoid using this equipment with high frequency electric knife high frequency therapy equipment or defibrillators and other electronic devices or may an electric shock occur to the patient Attention Using radio transmitting equipment nearby the system may interfere with the normal operation of the system Prohibited carry or use of devices that can generate radio wa...

Страница 11: ...onnect the grounding wire to any gas pipe or water pipe or it may cause bad grounding and danger of explosion 3 This equipment is not waterproof not use this equipment in place where liquid may into the interior of the equipment Never pour any liquid on the equipment otherwise there will be danger of electric shock or cause equipment damage If accidentally spill liquid on the equipment turn off th...

Страница 12: ...ery Service personnel or operator can replace the battery Before replacing the battery please contact Kaixin Company replace the battery under the guidance of Kaixin Company Please use the battery provided by Kaixin Company 1 Install the battery With four M3 6 pan head screws fix the battery on the lower case of main unit paste the Velcro on the buckle postion then put the buckle into the battery ...

Страница 13: ...check before and after operation Before and after ultrasonic diagnosis to check if there are any exceptionally on the surface of the probe or cable jacket such as peeling cracks bulge or if the acoustic lens is reliable cleaned or disinfected 5 6 Main unit check before and after operation 5 6 1 Inspection before start up Check the following items before starting the machine Warning 1 Adapter has n...

Страница 14: ... possible damage or blockage 9 Clean the field and environment 5 6 2 Inspection after start up Check the following items after starting the machine 1 No abnormal voice strange smell and overheating appear 2 Check the machine to ensure a normal start up The power indication light is on the machine will automatically enter B mode 3 Check the acoustic lens for abnormal heat when the probe is in use T...

Страница 15: ...System Set menu 2 Press or key to move the cursor to Time item and then press or key to adjust the time 3 Press key to confirm the time setting and quit system setting interface 6 2 3 Probe protect time setting 1 Press key and then press key to enter System Set menu 2 Press or key to move the cursor to Probe Time item and then press or key to adjust the probe protect time 3 Probe protect time is 0...

Страница 16: ...Press key to confirm this setting and quit system setting interface Explanation The ear tag scanning function is to be upgraded 6 2 7 Menu time setting 1 Press key and then press key to enter System Set menu 2 Press or key to move the cursor to Menu Time item and then press or key to adjust the menu disappear time 3 Menu disappear time is 5 30 seconds 4 Press key to confirm this setting and quit s...

Страница 17: ... Repeatedly press key also to achieve the cycle conversion among the B A 2B mode 6 4 Image adjustment 6 4 1 Image adjustment in B 2B mode In B 2B mode and in real time status you can adjust images gain dynamic range near field far field compression curve frequency framge average color brightness and contrast 6 4 1 1 Gain adjustment 1 Press key to enter image adjustment menu press or key to enter G...

Страница 18: ...press direction key to reduce far field gain so as to control the gain in far field region 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be saved and exit the image adjustment menu 6 4 1 5 Curve adjustment 1 Press key to enter image adjustment menu press or key to enter Curve item 2 Press or key to adjust the parameter within 0 3 3 After waitting for 5 30s Menu Tim...

Страница 19: ...de and in real time status press key to enter gain adjustment status 2 Press or key to adjust the gain within 27 90 3 After waitting for 5 30s Menu Time set in the System Set the parameters will be saved and exit the gain adjustment status 6 5 Depth range selection In B 2B mode and in real time status press or key to obtain the images of different depth It has three kinds of depth range selection ...

Страница 20: ...nuary 14 2016 on the storage path U IMAGE save a 20160114 csv EXCEL spreadsheet 6 7 1 2 Save data in A mode 1 Insert OTG U disk into the Micro USB interface at the right side of main unit the lower left corner of the screen appears U Disk prompt 2 Complete the A mode measurement freeze the image 3 Press key the system issued a tick tone namely the current data is stored in the EXCEL spreadsheet in...

Страница 21: ...ppropriate site of swine s back if necessary cut off the hair at the measurement site and wet the skin with hot water then apply light oil to wet the skin one minute or two in order to make good contact between swine skin and sound window The probe is placed against the skin of the swine and is kept perpendicular to the back move the probe gently make sure no air bubbles between the skin and probe...

Страница 22: ... make good contact between swine skin and sound window The probe is placed against the skin of the swine and is kept perpendicular to the back move the probe gently make sure no air bubbles between the skin and probe Keep the swine quiet during the test you can feed the swine during the test 3 Press key on the keyboard or press the freeze switch on the probe to unfreeze the image Press or key choo...

Страница 23: ... following table Glutaraldehyde based disinfectant Chemical Name Reagent Name Step Glutaraldehyde 2 4 Cidex Glutaraldehyde disinfectant Please refer to the instructions of the solution for details Warning Turn off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any...

Страница 24: ...You may follow the recommended method by the disinfectant manufacturer to rinse 4 After finishing the rinsing use a sterilized cloth or gauze to wipe the water on the surface of probe Do not dry the probe by heating it Immersion disinfection of probe sketch map 3 Check probe after cleaning and disinfection 1 Check the probe enclosure and its cable to ensure they are free of abnormity such as scuff...

Страница 25: ...ng will shorten the battery life the full charged battery can be used about 5 hours 4 Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use time reduced significantly compared with normal conditions the battery should be promptly replaced 5 The excess high or low temperature will affect the charging and discharging performance and short the...

Страница 26: ...ich will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off battery such as dry battery or different capability or different model or different brand battery which will lead to the leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or turn colors or abnormal phenomena which will lead to the leak...

Страница 27: ...far field Control panel malfunction Turn off the machine and restart it 8 4 Periodic Safety Checks To ensure the system performance and safety it must be checked after using 1 year When check the instrument please consult the International Trade Dept of Kaixin or its dealers as they need to have professional technology engineers 1 The following safety checks should be performed at least every 12 m...

Страница 28: ...and corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter 10 Safety Classification 1 Classified according to electric shock protection type Class I internally powered eq...

Страница 29: ...uzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Information contained in this manual is subject to change without further notice ...