Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet) User’s Manual V1.07
- 4 -
Chapter 4 Operation Condition
4.1 Power Requirements
Adapter rating: 100-240V~, 1.2-0.6A, 50-60Hz
Adapter model: BJE01-40-001M
Output of adapter: DC12.8V 3.0A
Main device rating: DC12V 3.0A
4.2 Operation Environment
Ambient temperature: 10
℃
-40
℃
Relative humidity: 30%-75% (without condensation)
Atmospheric pressure: 800hPa-1060hPa
4.3 Transport and Storage
Ambient temperature: -20
℃
-55
℃
Relative humidity: 30%-93% (without condensation)
Atmospheric pressure: 700hPa-1060hPa
Attention:
The main voltage is varies with different countries or regions.
Danger:
Do not use this equipment where flammable gas (such as anesthetic gas,
oxygen or hydrogen) or flammable liquid (such as alcohol) are present. Failure to do
so may result in explosion.
Warning:
Avoid using this equipment with high-frequency electric knife,
high-frequency therapy equipment or defibrillators and other electronic devices, or
may an electric shock occur to the patient.
Attention:
Using radio transmitting equipment nearby the system may interfere
with the normal operation of the system. Prohibited carry or use of devices that can
generate radio waves within the room installed this system, such as cell phones, radio
transceivers and wireless remote control toys.
Attention:
System should be avoided using in following environments:
1. Moist
2. Rain 3. Thunderstorm weather
4. No ventilation 5. Close to heat source
6. Direct sunlight
7. Dramatic temperature change
8. Poisonous gas
9. Corrosive gas
10. Strong shock
11. Strong electromagnetic field (e.g. MRI)
12. Radiation (e.g. X-ray, CT)
13. Defibrillators or short wave therapy equipment
Warning:
AC/DC adapter is as a part of the equipment, please only use the AC/DC
adapter provided by manufacturer.