Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments (Vet)                                                    User’s Manual V2.14 

 

- 8 -

1. The temperature, humidity and atmospheric pressure shall meet the requirements of operation 

condition. 

2. No condensation occurs. 
3. No distortion, damage or contamination on system and peripheral. Clean the parts as specified in 

relevant sections, if the contaminant is present. 

4. Check the control panel, LED screen and enclosure to ensure they are in good working condition and 

free of abnormity (such as cracks and loosened screws). 

5. No damage on power cable, and hard up on its connection. 
6. Check probe and its connections to ensure they are free of abnormity (such as scuffing, drop-off or 

contamination). If the contaminant is present, clean, disinfect the contaminated objects as specified 
in relevant sections. 

7. No barriers around the intake of equipment. 
8. See to it that probe has been cleaned, disinfected; else dispose it as specified in relevant sections. 
9. Check all the ports of the machine for possible damage or blockage. 
10. Clean the field and environment. 

5.8.2 Inspection after start-up   

Check the following items after starting the machine: 

1. No abnormal voice, strange smell and overheating appear. 
2. Check the machine to ensure a normal start-up: The power indication light is on and startup picture is 

shown on the screen. Then the machine will be automatically set in B mode. 

3. Check the image to ensure trouble-free display (no excessive noise or flicker). 
4. Check the control panel to ensure normal operation condition. 
5. Check the instrument to ensure that the phenomenon of local high temperature will not appear.   

 
 
 
 

 

5.9 System reset 

In case of abnormal screen display or no-working for system operation, try to restart the system by 

turning on/off the main unit power. 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Attention

:  Thoroughly  clean  the  coupling  gel  on  the  probe  surface  each  time  after 

ultrasonic operation, or the coupling gel may become hardened on the acoustic lens of the 
probe, deteriorating quality of image. 

Содержание 3256

Страница 1: ...KAIXIN MSU2 Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd ...

Страница 2: ...f this user s manual without the express written permission of Kaixin Statement Kaixin has the final explanation right of this user s manual Kaixin was considered responsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement parts...

Страница 3: ...ems damages or losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic process Kaixin shall not be liable for any problems arising out of the process 7 Be sure to back up important data to external storage media such as notebooks ...

Страница 4: ...ial danger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided Indicates that make sure refer to relevant contents in this manual 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Keep dry Power supply indication Fragile Direct current Stacking limit by number Follow instructions for use Temperature limits Up Humidity ...

Страница 5: ...n 7 5 8 Main unit check before and after operation 7 5 9 System reset 8 Chapter Six Functional Operation 9 6 1 Startup and Shutdown 9 6 2 Screen display reverse 9 6 3 System Functions Setting 9 6 3 1 Time Setting 9 6 3 2 TV Mode Setting 9 6 3 3 Energy Saving Setting 9 6 3 4 Characters Brightness Setting 9 6 3 5 Hospital Name Setting 9 6 3 6 Key Sound Setting 10 6 3 7 Chinese English Setting 10 6 4...

Страница 6: ...10 Delete the image 14 6 11 Text Input 14 6 12 Obstetric list management 14 6 12 1 Save and view obstetric list 14 6 12 2 Delete obstetric lists 15 Chapter Seven General Measurement 16 7 1 Distance Measurement 16 7 2 Circumference Area Volume Measurement 16 7 3 Slope Heart rate Cycle Measurement 17 Chapter Eight Obstetric Measurement 18 8 1 Measurement and Calculation items 18 8 2 Measurement of G...

Страница 7: ...e correlation function 5 Image post process function 6 Image freeze unfreeze 7 Depth range selection function 8 Color display function 9 Image filter function 10 Automatic backfat measure function 11 Local zooming 12 Adjustment and display of near field far field total gain and dynamic range 13 Image storage 14 Image vertical horizontal reverse 15 Case information image annotation and auto time di...

Страница 8: ...arts of the probe Fig Parts name of 3 5MHz mechanical sector probe Name Function 1 Acoustic lens Use mechanical methods so that the sound beam transmitted by the transducer can sector scanning for a certain angle 2 Cable To connect the probe to the probe connector 3 Probe connector To connect the probe to ultrasonic diagnostic system 4 Freeze switch Press the Freeze switch to freeze unfreeze the i...

Страница 9: ...mperature 10 40 Relative humidity 30 75 without condensation Atmospheric pressure 700hPa 1060hPa 4 3 Storage and transport Ambient temperature 20 55 Relative humidity 30 93 without condensation Atmospheric pressure 700hPa 1060hPa Danger Do not use this equipment where flammable gas such as anesthetic gas oxygen or hydrogen or flammable liquid such as alcohol are present Failure to do so may result...

Страница 10: ...ment the case may damage the instrument 7 The instrument is turned off completely only by disconnecting the power supply from the wall socket Warning 1 If breakers and fuse of the mains power socket are identical to those of this system and they are used to control the current for equipment like life support system the system shall not be connected to such power supply socket as it may cause break...

Страница 11: ...d 3 Fig Install the hood 2 The installed hood should be inserted the Hood buckle position aligned with the side of the instrument and then be inserted into the other side Note The force should be uniform 3 Take out the attached strap install it on the the hook of hood or put it on the the hole of strap of the instrument as shown in the figure below 4 Adjust the length of strap hanging on the neck ...

Страница 12: ...c shock Turn off the system and use socket screwdriver to remove the probe with figure s method and instructions Warning 1 The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic system only Select proper probe model according to the relevant instructions of ultrasonic diagnostic system 2 Check the ultrasonic probe and connecting cable after diagnostic operation Use of de...

Страница 13: ...power adapter Insert the output plug of adapter into DC power input interface 2 Connect to the main power supply Insert the power plug jack furnished with the machine into power input socket of the power adapter the other end to the mains socket outlet The instrument uses three core power supply It connects with the protective earth line when power plug inserts into its socket 5 7 Ultrasonic probe...

Страница 14: ...aned disinfected else dispose it as specified in relevant sections 9 Check all the ports of the machine for possible damage or blockage 10 Clean the field and environment 5 8 2 Inspection after start up Check the following items after starting the machine 1 No abnormal voice strange smell and overheating appear 2 Check the machine to ensure a normal start up The power indication light is on and st...

Страница 15: ...etting interface 6 3 2 TV Mode Setting 1 Freeze image press direction keys to move cursor to Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to point to TV mode 4 Press direction keys to achieve TV mode conversion between PAL and NTSC 5 Press key to quit setting interface 6 3 3 Energy Saving Setting 1 Freeze image press direction keys to move cursor t...

Страница 16: ...to move cursor to Preset 2 Press direction keys to enter setting interface 3 Press key to switch between Chinese and English 4 Press key to quit setting interface 6 4 Mode Selection In real time mode press key repeatedly to achieve mode switching of B B B 4B B M and M 6 4 1 B Mode B mode is a basic operation mode after startup and a single framed B mode image is displayed 6 4 2 B B Mode 1 In real ...

Страница 17: ... the display area Press direction key to increase near field gain and direction key to reduce near field gain so as to control the gain in near field region 6 5 4 Far Field Gain Adjustment In real time mode press key to enter main menu and press direction keys to move cursor to Far in the display area Press direction key to increase far field gain and direction key to reduce far field gain so as t...

Страница 18: ...Selection In real time mode press key to enter main menu and press direction keys to move cursor to Color in the display area and then press direction keys to achieve the conversion of eight kinds of colors including one kind of black and white 6 6 7 Image Filter In real time B mode press key to enter main menu and press direction keys to move cursor to Filter in the display area and then press di...

Страница 19: ... the depth Press key to quit 6 8 Body Mark and Probe Mark This product contains 27 body marks that are divided into two pages when display The operation steps are as follows 1 Freeze the image press direction keys to move cursor to BodyMark press key body marks will be showed in the image area press direction keys to change pages 2 Press direction keys to move to the position of desired body mark ...

Страница 20: ...ft 5 6 7 8 9 N O P Q R S T U V W X Y Z Press direction keys to move cursor to point to Caps and then press key to achieve capital and small letter conversion If the cursor point to Shift press key to achieve the conversion between the letter and punctuation 4 Press direction keys to choose numbers or characters and press key to confirm 5 After inputting EAR ID press key the screen below shows Preg...

Страница 21: ...der is automatically named by the system current date the EXCEL spreadsheet is automatically named by the system current time 6 Press direction keys to view stored obstetric lists 7 Press direction keys to exit obstetric list interface and press key to return to real time status Such as U 032217 1101 csv It represents that on March 22 2017 on the storage path U 032217 save an 1101 EXCEL spreadshee...

Страница 22: ... clear all marks and data and quit the measurement status 7 2 Circumference Area Volume Measurement Circumference area volume measurement with ellipse method 1 In B B B mode freeze desired image the cursor is located in the Measure position of display area 2 Press key the measurement methods are showed in the lower left of the screen press direction keys to choose Ellipse press key again the curso...

Страница 23: ...mplete the multi group data measurement 7 Continuously press key twice to quit the measurement status 8 Press direction keys to choose Clear and then press key to clear all marks and data 9 Press key to unfreeze clear all marks and data and quit the measurement status 7 3 Slope Heart rate Cycle Measurement The method to measure slope hear rate cycle is identical with distance measurement Note In B...

Страница 24: ...n as the starting point of the measurement 5 Press direction keys to move the mark to the end point of the measurement At the same time a lighted dotted line appears between the start and the end as the dashed locus of the measurement The measured value is automatically displayed at the built in mark mm on the right side of the screen G Aand EDC value to be displayed in real time in the right area...

Страница 25: ...g Turn off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any waterproof device in the system Do not splash any water or liquor into the system when cleaning or maintaining otherwise it will cause malfunction or electric shock Attention 1 Probe without cleaning di...

Страница 26: ... least one minute to remove the residual chemicals You may follow the recommended method by the disinfectant manufacturer to rinse 4 After finishing the rinsing use a sterilized cloth or gauze to wipe the water on the surface of probe Do not dry the probe by heating it Fig Wrong immersion disinfection of 3 5MHz probe Fig Right immersion disinfection of 3 5MHz probe 3 Check probe after cleaning and...

Страница 27: ...an be used for 5 6 hours 3 Battery is consumable the battery cycle life is based on the times of charge and discharge as unit When the use time reduced significantly compared with normal conditions the battery should be promptly replaced 4 The excess high or low temperature will affect the charging and discharging performance and short the battery life and capacity Attention Battery is consumable ...

Страница 28: ...Do not disassemble the battery in any way which will lead to the leakage explosion or fire 10 Do not charge the battery near the heat source or extra hot environment which will lead to the leakage explosion or fire 11 Do not put the battery into the microwave oven or pressure vessel which will lead to the leakage explosion or fire 12 Do not mixed use the battery together with one off battery such ...

Страница 29: ...t of Kaixin or its dealers as they need to have professional technology engineers 1 The following safety checks should be performed at least every 12 months by a qualified person who has adequate training knowledge and practical experience to perform these tests Inspect the equipment and accessories for mechanical and functional damage Inspect the safety relevant labels for legibility Inspect the ...

Страница 30: ...and corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shock protection type Class I internally powere...

Страница 31: ...uzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Information contained in this manual is subject to change without further notice ...

Отзывы: