Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet)                                                    User’s Manual V1.03 

 

- 22 - 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 
 
 
 
 
 
 
 

9.2 Troubleshooting 

To  ensure  normal  operation,  users  are  recommended  to  prepare  a  proper  maintenance  and  regular 

examination plan to regularly check on product safety performance. If any abnormity occur, timely contact 
International Trade Dept of Kaixin for support. 

If  the  following  problems  occur  on  starting  up  the  machine,  try  to  make  corrections  following  the 

method  in  the  table.  If  the  problem  remains  unsolved,  contact  International  Trade  Dept  of  Kaixin  for 
support. 

Trouble 

Correction 

Power light is off and no screen 
display is present when starting the 
machine. 

1. Check power supply. 
2. Check power cable and connector. 
3. Check power adapter. 

Attention: 

1.

 

Do  not  throw  the  battery  into  water  or  be  wet,  which  will  lead  to  the  battery  leakage, 
explosion or fire; 

2.

 

Do not use or store the battery near the heat source, such as fire or heater, which will lead 
to the battery leakage, explosion or fire; 

3.

 

Do  not  connect  the  anode  and  cathode  reversely,  which  will  lead  to  the  battery  leakage, 
explosion or fire; 

4.

 

Do not heat up or throw the battery into fire, which will lead to the leakage, explosion or fire; 

5.

 

Do not connect the anode and cathode with any metal or conductor; do not transport or 
store the battery together with necklaces, hairpins or other metal objects, which will lead 
to the leakage, explosion or fire; 

6.

 

Do  not  hammerblow,  throw  or  mechanically  shake  the  battery,  which  will  lead  to  the 
leakage, explosion or fire; 

7.

 

Do  not  insert  the  battery  with  nail  or  other  spiculate  objects;  do  not  hammerblow  or 
trample the battery, which will lead to the leakage, explosion or fire; 

8.

 

Do not weld the battery terminal directly, which will lead to the leakage, explosion or fire; 

9.

 

Do not disassemble the battery in any way, which will lead to the leakage, explosion or fire; 

10.

 

Do not charge the battery near the heat source or extra-hot environment, which will lead 
to the leakage, explosion or fire; 

11.

 

Do not put the battery into the microwave oven or pressure vessel, which will lead to the 
leakage, explosion or fire; 

12.

 

Do  not  mixed  use  the  battery  together  with  one-off  battery  (such  as  dry  battery),  or 
different  capability  or  different  model  or  different  brand  battery,  which  will  lead  to  the 
leakage, explosion or fire; 

13.

 

Do not use the abnormal battery with particular smell or abnormal heat or distortion or 
turn colors or abnormal phenomena, which will lead to the leakage, explosion or fire; 

14.

 

Do  stop  the  charge  and  pull  out  the  battery  from  the  charger  at  once  if  any  abnormal 
phenomenon happens to the battery, such as particular smell or abnormal heat or distortion 
or turn colors. Otherwise, each of above will lead to the leakage, explosion or fire; 

15.

 

Do remove the battery from the near fire if any leakage or particular smell happens, which 
will lead to the leakage, explosion or fire; 

16.

 

If any leakage splash into eye, do not wipe the eye, instead of washing it and get help from 
the doctor as soon as possible. Otherwise, the eye will be injured; 

17.

 

Do not use the battery in the extremely hot environment, such as hot sunshine or in the car 
when it is too hot, because these will catch fire, even worsen its performance and shorten 
its life; 

18.

 

If  use  the  battery  beyond  the  listed  environment  on  the  manual,  it  will  worsen  its 
performance or shorten its life, even lead to extreme heat or explosion or fire. 

Содержание 2857

Страница 1: ...KAIXIN MSU1Plus Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments Vet User s Manual Xuzhou Kaixin Electronic Instrument Co Ltd ...

Страница 2: ...rt of this user s manual without the express written permission of Kaixin Statement Kaixin has the final explanation right of this user s manual Kaixin was considered responsible for the safety reliability and performance in case of meeting all the following requirements 1 Assembly expansion readjustment improve and repair are all performed by professionals recognized by Kaixin 2 All replacement p...

Страница 3: ...ems damages or losses due to re installation alteration or repair the system by non Kaixin designated personnel 6 This product is intended to provide clinical diagnostic data for the doctor The doctor shall be responsible for the diagnostic process Kaixin shall not be liable for any problems arising out of the process 7 Be sure to back up important data to external storage media such as notebooks ...

Страница 4: ...ot avoided Attention Indicates a potential danger or unexpected use condition that may result in light injury or property loss or affecting the use if not avoided Indicates that make sure refer to relevant contents in this manual 2 Safety Symbols Symbols Meaning Symbols Meaning Type B applied part Fragile Direct current Stacking limit by number Power supply indication Temperature limits Follow ins...

Страница 5: ... 7 5 6 Connect to power 7 5 7 Ultrasonic probe check before and after operation 8 5 8 Main unit check before and after operation 8 5 9 System reset 8 Chapter Six Functional Operation 9 6 1 Startup and Shutdown 9 6 2 System Functions Setting 9 6 2 1 Time Setting 9 6 2 2 TV Mode Setting 9 6 2 3 Energy Saving Setting 9 6 2 4 Characters Brightness Setting 9 6 2 5 Hospital Name Setting 9 6 2 6 Key Soun...

Страница 6: ... recall 13 6 9 1 Save image 13 6 9 2 Open image 14 6 10 Delete the image 14 6 11 Obstetric records management 14 6 11 1 Save and view obstetric records 14 6 11 2 Delete obstetric records 15 Chapter Seven General Measurement 16 7 1 Distance Measurement 16 7 2 Circumference Area Volume Measurement 16 7 3 Slope Heart rate Cycle Measurement 17 Chapter Eight Obstetric Measurement 18 8 1 Measurement and...

Страница 7: ...tion 5 Image post process function 6 Image freeze unfreeze function 7 Depth range selection function 8 Pseudo color display function 9 Image filter function 10 Backfat ruler display function 10 Local zooming function 11 Adjustment of near field far field total gain and dynamic range 12 Image storage function 13 Image up down left right reverses function 14 Case information image annotation and aut...

Страница 8: ...me of 3 5MHz mechanical sector probe Name Function 1 Acoustic lens Use mechanical methods so that the sound beam transmitted bythe transducer can sector scanning for a certain angle 2 Cable To connect the probe to the probe connector 3 Probe connector To connect the probe to ultrasonic diagnostic system 4 Freeze switch Press the Freeze switch to freeze unfreeze the image 2 4 Function keys descript...

Страница 9: ...ative humidity 30 75 without condensation Atmospheric pressure 700hPa 1060hPa 4 3 Storage and Transport Ambient temperature 20 55 Relative humidity 30 93 without condensation Atmospheric pressure 700hPa 1060hPa Attention The mains voltage is varies with different countries or regions Attention System should be avoided using in following environments 1 Moist 2 Rain 3 Thunderstorm weather 4 No venti...

Страница 10: ...cessing equipment and IEC 60601 1 for medical equipment If in doubt consult the technical service department or your local representative 4 When this system is installed or used in the patient environment try to avoid the patient touching the system If system is with some unknown defects it may cause danger of electric shock 5 If the integrity of the external protective conductor in the installati...

Страница 11: ...main unit as fig shown b Adjust the strap length the front strap hung on the neck the back strap fixed on the waist two straps can also adjust to similar length to meet the requirements of the screen visual angle while hanging around the neck to use Installation a Take out the instrument and accessory install the probe b The probe passes through the oval hole of the leather bag and put the instrum...

Страница 12: ...lectric shock Warning 1 The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic system only Select proper probe model according to the relevant instructions of ultrasonic diagnostic system 2 Check the ultrasonic probe and connecting cable after diagnostic operation Use of defective probe may cause electric shock 3 Do not knock or bump the probe or it may be damaged and ca...

Страница 13: ... the power input socket of the video recorder the other end to the power supply socket 5 6 Connect to power 1 Connect to the power adapter Insert the output plug of adapter into power input interface of main unit 2 Connect to the main power supply Insert the power plug jack furnished with the machine into power input socket of the power adapter the other end to the main power supply socket The ins...

Страница 14: ...the contaminant is present clean disinfect the contaminated objects as specified in relevant sections 7 No barriers around the intake of equipment 8 See to it that probe has been cleaned disinfected else dispose it as specified in relevant sections 9 Check all the ports of the machine for possible damage or blockage 10 Clean the field and environment 5 8 2 Inspection after start up Check the follo...

Страница 15: ...ress direction keys to enter setting interface 3 Press direction keys to move symbol to point to TV mode 4 Press direction keys to achieve TV mode conversion between PAL and NTSC 5 Press key to quit setting interface 6 2 3 Energy Saving Setting 1 Freeze image press direction keys to move cursor to Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to poi...

Страница 16: ...face 3 Press key to switch between Chinese and English 4 Press key to quit setting interface 6 3 Mode Selection In real time mode press key repeatedly to achieve mode switching of B B B 4B B M and M 6 3 1 B Mode B mode is a basic operation mode after startup and a single framed B mode image is displayed 6 3 2 B B Mode 1 In real time mode press key to enter B B mode 2 B B image switch Press key to ...

Страница 17: ...ction key to increase near field gain and direction key to reduce near field gain so as to control the gain in near field region 6 4 4 Far Field Gain Adjustment In real time mode press key to enter main menu and press direction keys to move cursor to Far in the display area Press direction key to increase far field gain and direction key to reduce far field gain so as to control the gain in far fi...

Страница 18: ...enu and press direction keys to move cursor to Color in the display area and then press direction keys to achieve color conversion of eight colors including one kind of black and white 6 5 7 Image Filter In real time B mode press key to enter main menu and press direction keys to move cursor to Filter in the display area and then press direction keys to achieve the conversion of two kinds of filte...

Страница 19: ...ocated behind ID 3 Press key again at the same time characters input menu will be shown at the bottom of the screen Caps 0 1 2 3 4 AB C D E F G H I J K L M Shift 5 6 7 8 9 N O P Q R S T U V W X Y Z Press direction keys to move cursor to point to Caps and then press key to achieve capital and small letter conversion If the cursor point to Shift press key to achieve the conversion between the letter...

Страница 20: ...se the image code with mark 6 10 Delete the image 1 Freeze the image press key to enter reading image status 2 Press direction key to select the image code to be deleted such as 002 3 Press key the image will be deleted will automatically disappear 4 Repeat the 2 3 steps delete other images 5 Press key to return to real time mode 6 11 Obstetric records management 6 11 1 Save and view obstetric rec...

Страница 21: ... 1511 csv It represents that on October 18 2016 on the storage path U 161018 save a 1511 csv EXCEL spreadsheet Note 1 The EXCEL spreadsheet records the results for each inspection in the order In the EXCEL spreadsheet the contents of each column respectively ID Result Time and Remark Note 2 The main unit stores up to 256 obstetric records 6 11 2 Delete obstetric records 1 Freeze B mode image press...

Страница 22: ...n B mode freeze desired image press direction keys to move cursor to Measure or in the B B mode freeze desired image the cursor is located in the Measure position of display area 2 Press key the measurement methods are showed in the lower left of the screen press direction keys to choose Ellipse press key again the cursor will show 3 Press direction keys to move the mark to desired position press ...

Страница 23: ...eys to choose Clear and then press key to clear all marks and data 9 Press key to unfreeze clear all marks and data and quit the measurement status 7 3 Slope Heart rate Cycle Measurement The method to measure slope hear rate cycle is identical with distance measurement Note In B M mode if both starting point and end point of the measurement mark fall into the B mode area the value of the refers to...

Страница 24: ...urement 5 Press direction keys to move the mark to the end point of the measurement At the same time a lighted dotted line appears between the starting point and the end point as the dashed locus of the measurement The measured value is automatically displayed at the built in mark mm on the right side of the screen G A and EDC value to be displayed in real time in the right area of the screen 6 Pr...

Страница 25: ...tep Glutaraldehyde 2 4 Cidex Glutaraldehyde disinfectant Please refer to the instructions of the solution for details Warning Turn off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any waterproof device in the system Do not splash any water or liquor into the sys...

Страница 26: ...ed cloth or gauze to wipe the water on the surface of probe Do not dry the probe by heating it Fig Wrong immersion disinfection of 3 5MHz probe Fig Right immersion disinfection of 3 5MHz probe 3 Check probe after cleaning and disinfection 1 Check the probe enclosure and its cable to ensure they are free of abnormity such as scuffing cracks or drop off 2 The sound window of probe is thin ensure tha...

Страница 27: ...normal conditions the battery should be promptly replaced 4 The excess high or low temperature will affect the charging and discharging performance and short the battery life and capacity Attention A power indicator will appear X and glitter continually when the electric quantity is too low Connect the main unit to external power supply and recharge the battery or turn off the machine to recharge ...

Страница 28: ...erblow or trample the battery which will lead to the leakage explosion or fire 8 Do not weld the battery terminal directly which will lead to the leakage explosion or fire 9 Do not disassemble the battery in any way which will lead to the leakage explosion or fire 10 Do not charge the battery near the heat source or extra hot environment which will lead to the leakage explosion or fire 11 Do not p...

Страница 29: ...fessional technology engineers 1 The following safety checks should be performed at least every 12 months by a qualified person who has adequate training knowledge and practical experience to perform these tests Inspect the equipment and accessories for mechanical and functional damage Inspect the safety relevant labels for legibility Inspect the fuse to verify compliance with rated current and br...

Страница 30: ...and corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shock protection type Class I internally powere...

Страница 31: ...uzhou Jiangsu China Post Code 221004 Tel 86 516 87732932 87733758 Fax 86 516 87732932 87792848 Website http www kxele com E mail info kxele com Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Information contained in this manual is subject to change without further notice ...

Отзывы: