Full digital mechanical sector scanning ultrasonic diagnostic instruments (Vet) User’s Manual V1.03
- 6 -
5.2 Ultrasonic probe installation
5.2.1 Ultrasonic probe connection
Insert the probe connector into the probe socket at the back of the main unit.
5.2.2 Ultrasonic probe disconnection
Socket screwdriver
Button position on both side of socket
According to arrow
direction,gently insert the socket
screwdriver, then button up the
button position on both side of the
socket with your fingers,you can
remove the probe socket.
5.3 Install/Remove the battery
1. Install the battery
Put the battery into battery storage of main unit; cover the “Upper cover of battery storage” into the
corresponding position, and then toggle the push button in the opposite direction, lock the “Upper cover
of battery storage”.
2. Remove the battery
According to the direction of the arrow, toggle “Push button of battery storage”, removed “Upper
cover of battery storage” as shown in figure, take out the battery.
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
probe may cause electric shock.
Warning:
1. The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic system
only. Select proper probe model according to the relevant instructions of ultrasonic
diagnostic system.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not knock or bump the probe, or it may be damaged and cause electric shock.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
2. Before disconnecting the probe, place the cable and probe on a stable and leveled position
so that the probe may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of service
life of probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.
Turn off the system, and use
socket screwdriver to remove the
probe with figure’s method and
instructions.
Danger:
Use together with flammable anaesthetic, it may result in explosion.
Attention:
Probe is highly sensitive to shake, be used with caution. About probe’s use
and cleaning, the details see the relevant sections.