HygroPort I.S. User’s Manual
Table C.1 shows the status flags associated with each Bit of the status register. If a
status fl lag is set high then the associated condition exists. If it is set low, then that
condition does not exist.
Details
Reading out of range (beyond
+30/-120ºC)
Reading out of range (beyond
+100/-40ºC)
External Sensor Input out of range
(<4 mA / >20 mA)
Battery voltage <4.7 V
Internal dewpoint not stable
Internal dewpoint stable
Log file has reached the maximum
of 10,000 logs and logging has
been stopped
Could not detect internal sensor on
power up
Config. fi ile missing, new file
created and default settings used
Indicates system in purging phase
Table C.1
Status register flags
For the example given in Figure C.1, from Table C.1, Hex 28 translates, reading from
right to left, as Bit 4 set (
External sensor error
) and Bit 6 set (
Measurement in
progress
). Therefore, at the time that these log points were taken, the external sensor
was out of range and the internal dewpoint was not stable.
Issue 6, December 2019
56
Kahn Instruments
Status Register
Bit in Hex
0x0002 = BIT 2
0x0004= BIT 3
0x0008 = BIT 4
0x0010 = BIT 5
0x0020 = BIT 6
0x0040 = BIT 7
0x0080 = BIT 8
0x0100 = BIT 9
0x0200 = BIT 10
0x0800 = BIT 12
Message
Internal sensor error
Internal thermistor
error
External sensor error
Battery low
Measurement in
progress
Measurement OK
Log finished
Internal sensor not
found
Default CONFIG file
Used
Initializing internal
sensor
Содержание HygroPort I.S.
Страница 2: ...Inside front cover blank...
Страница 36: ...HygroPort I S User s Manual 3 4 5 Accessories Issue 6 December 2019 25 Kahn Instruments 1 6 2 7 Figure 3 3...
Страница 59: ...HygroPort I S User s Manual Appendix A Technical Specifications Issue 6 December 2019 48 Kahn Instruments...
Страница 62: ...HygroPort I S User s Manual Appendix B Hazardous Area Certification Issue 6 December 2019 51 Kahn Instruments...
Страница 65: ...HygroPort I S User s Manual Appendix C Datalog Status Display Issue 6 December 2019 54 Kahn Instruments...
Страница 68: ...HygroPort I S User s Manual Appendix D FCC Declaration Issue 6 December 2019 59 Kahn Instruments...
Страница 70: ......