INSTALLATION
2.5.1 Transmitter Mounting - Sample Block (Optional)
The following procedure must be carried out by a qualified
installation engineer.
To mount the transmitter into the sensor block (preferred method), proceed as follows,
refer to
Figure 12.
1.
Insure that the green, blue or black protective cover (2), and its
desiccant capsule (2a), have been removed from the tip of the
transmitter.
2.
G 1/2” and 5/8” Versions - Insure that the bonded seal (2) is over the
threaded part of the transmitter body.
3/4” Version - Insure the O-ring is fully seated in the recess.
Under no circumstances should the sensor guard be handled
with the fingers.
3.
Screw the transmitter (1) into the sample block (3) and tighten to the
appropriate torque setting (see Section 2.5).
NOTE: Use the flats of
the hexagonal nut and not the sensor body.
4.
Fit the transmitter cable/connector assembly to the plug located on the
base of the transmitter and tighten the securing screw (see Section 2.3).
1
2
3
5
7
6
2a
4
4
Figure 12
Transmitter Mounting
Содержание Easidew I.S.
Страница 19: ...Easidew I S User s Manual Kahn Instruments 13 APPENDIX A Appendix A Technical Specifications...
Страница 22: ...Easidew I S User s Manual 16 97099 Issue 13 4 August 2015 APPENDIX B Appendix B System Drawings...
Страница 26: ...Easidew I S User s Manual 20 97099 Issue 13 4 August 2015 APPENDIX C Appendix C Hazardous Area Certification...
Страница 29: ...Easidew I S User s Manual Kahn Instruments 23 APPENDIX D Appendix D EC Declaration of Conformity...
Страница 30: ...Easidew I S User s Manual 24 97099 Issue 13 4 August 2015 APPENDIX D Appendix D EC Declaration of Conformity...
Страница 31: ...http www kahn com...