background image

 

 

 

 
 
 
 

K-SYSTEMS 

USER MANUAL 

 
 

Anti Vibration Table - AV1 and AV5 

 

Version 05 

 
 
 

  

 

 

 

Manual part no: 33095 

Содержание AV1

Страница 1: ...1 K SYSTEMS USER MANUAL Anti Vibration Table AV1 and AV5 Version 05 Manual part no 33095 ...

Страница 2: ...ssembling the AV1 AV5 4 3 1 Weight adjustment 6 3 1 1 Removing the middle spring 7 3 1 2 Adding the Weight Plate 8 4 Trouble Shooting 9 5 Maintenance 10 6 Technical data AV1 10 7 Warranty Liability issues 11 7 1 Limited Warranty 11 7 2 Liability 11 7 3 Replacement 12 8 Technical assistance 12 ...

Страница 3: ...on The K SYSTEMS Anti Vibration tables are designed to meet the requirements of ICSI Procedures With an isolated microscope platform and a complex anti vibration system the table is preventing vibrations from reaching the microscope platform and dampening unavoidable vibrations in the shortest possible time The inverted microscope is placed on an individually suspended floating platform The platfo...

Страница 4: ...TE The table can be modified to optimally work with equipment of different weight It is important that you know the exact weight of the equipment before proceeding See chapter 3 1 Put on the gloves Remove the foam pads from the springs without touching the springs with your bare hands Place the floating plate upon the springs leave equal distance from all sides to the hole of the table plate for c...

Страница 5: ...AV1 User Manual 5 The position and height of the floating plate are adjusted by Loosen the four securing bolts Turn the adjustment screws to the necessary position Tighten the four securing bolts again ...

Страница 6: ...s heavy equipment on the platform Step 1 To work with equipment from 30 45 kg one spring is removed so only 2 springs are used NOTE Only remove the middle spring when needed Step 2 To work optimal with equipment from 15 30 kg 2 springs are installed and an extra weight plate must be used 2 springs and Weight plate 2 Springs without Weight plate 3 Springs without Weight plate From factory Equipment...

Страница 7: ...V1 3 Please put on gloves when touching the springs 4 In each side of the table there are metal rods screwed into the frame unscrew these two metal rods 5 Push the springs to one end and lift out the other metal rod 6 Remove the screw in the middle spring on the opposite side of the end that is lifted 7 Pull out the middle spring 8 Push the two remaining springs to the bottom side 9 Move the other...

Страница 8: ...m the bottom plate remove the plate 3 Add the extra weight 4 Mount and screw the bottom plate on again AV5 table 1 Place the floater on a flat surface and remove the screws 2 Remove the bottom plate and add the extra weight 3 The extra weight in place 4 Mount and screw the bottom plate and reinsert screws ...

Страница 9: ...tion is lost and vibrations are transferred to the microscope Clean the areas with 70 alcohol Tight cables from the micromanipulators will cause vibrations to be transmitted directly from the micromanipulator to the microscope Loosen the cables so they hang loose The floater is placed too high resulting in the springs touching the underside of the table plate Lower the floater so it is in level wi...

Страница 10: ...dled carefully Be sure to keep these chemicals away from open flames or potential sources of ignition for example electrical equipment that is being switched on or off Also remember to always use these chemicals only in a well ventilated room 2 If smudges are difficult to clean wipe them with a soft cloth slightly moistened with a diluted neutral detergent 6 Technical data AV1 Material Stainless s...

Страница 11: ...r implied In particular K Systems Kivex Biotec does not warrant that the product is suitable for the needs of the purchaser and there are no warranties given as to merchantability or fitness for a particular purpose other than the one specified in K Systems Kivex Biotec literature that accompanies every specific product K Systems Kivex Biotec assumes that the Purchaser is experienced in the use of...

Страница 12: ...ance and instructions Only for all customers in Denmark Norway and Sweden please take direct contact with our head office at Tel 45 4599 5600 Fax 45 4599 5619 email info k systems dk 8 Technical assistance For more information contact the K systems office nearest you For Denmark call 45 45995600 Outside Denmark see your K systems catalogue for the phone number of the office nearest you or visit ou...

Отзывы: