Calibration
5.3 Saving Calibration Constants
Factory-calibrated constants are permanently stored in a bank of
the onboard EEPROM and may not be modified. When you re-cal
-
ibrate the device, the software stores the new constants in a user-
configurable section of the EEPROM. To return a device to its ini
-
tial factory calibration settings, the software can copy the factory
calibrated constants to the user-configurable section of the
EEPROM. When auto-calibration is completed, you can save the
new calibration constants to the user-configurable banks in the
EEPROM. The date and the temperature when you ran auto-cali
-
bration is saved with the calibration constants. You can store three
sets of calibration constants according to three different environ
-
ments and re-load the calibration constants later.
NOTES
• Warm the card up for at least 15 minutes before initiating auto-cal
-
ibration.
• Remove the cable before auto-calibrating the card since the DA
outputs are changed during the process.