Users Manual

Applicable Model: 09401

Two-channel Oscilloscope DSO 094

Panel & Connectors

Modes & Menu

DN094-02v01/2012-05

Basic Operations

1. Powering ON/OFF the Device

2. Setting Parameters

1 ) Unscrew the two screws at back cover. Remove back cover.

Battery was disconnected for transportation reason. You need
connect it before powering the device up. Please follow th e steps
below to c onnect batter y:

Channel conf ig.

Test Signal
1Vpp / 1KHz

Ch 1 (Y) Input

Ch 2 (X) Input

Ch 1 Couple
Sel ect

Ch 2 Couple
Select

Push: power ON
Hold: power OFF
Turn: adjustment

Tog gle RUN/HOLD

Ch 1 (Y) Sensitiviy & Position

Ch 2 (X) Sensitiviy & Position

Time base & Horizont al position

Trigger Lev el, Slope, Source, & Mode

Menu

Alternative functions

Test Signal
1Vpp / 1KHz

Ch 1 (Y) Input

Ch 2 (X) Input

USB Connect or

No connection

Ch 1
V. Pos

Ch 2
V. Pos

Trig Point
Indicator

Dis playe d portion
of capture buffer

Trigger level in dicator

Ch 1 sensitivity,
couple, & position

Ch 2 sensitivity,
couple, & position

Ti me ba se & H. Position

Trigger level,
slope, & source

Tri gger mode

Y Pos.

X Pos.

Y Pos.

X Pos.

Sampling
Rate

Y/T Mode

This is the most commonly used mode that
displays signal-to-time relationship.

Y/X Mode

This mode can be used to graph relationships
between two signals. In this mode Ch 1 becomes Y
and Ch 2 becomes X (see photo at right).

MENU

Menu Function Summary

Item

DSO Y/T

DSO Y/X

SAVE WAVEFORM

RECALL WAVEFORM

SEND SCREEN

Menu is used to switch modes and perform special functions. Press [MENU] botton to enter menu
and press [HOLD] to execute selected item. Press [MENU] to exit.

CHANGE REC. LEN

CHANGE TRIG POS

CHANGE TRIG SEN

RESTORE DEFAULT

REBOOT

Functions

Select Y/T mode

Select Y/X mode

Save captured waveform to EEPROM

Recall saved waveforms

Send screen to PC as bitmap file

Change record length (waveform buffer size)

Change trig position (pre-trig buffer size)

Change trigger sensitivity

Restore settings to factory default

Reboot the oscilloscope (enter bootloader)

Y/X Mode Screen

First Time Powering UP

2 ) Battery was attached to the inner side of back cover. Insert its

connector to the header on PCB (marked as "BATT" at the side of
USB connector) as shown in the photo at right.

3 ) Put back cover back and screw it up.

Power On

- presssing [ADJ] knob

Power Off

- holding down [ADJ] knob for about 2 - 3 seconds

Frequently operated paramete rs are grouped into four catagor ies and governed by four buttons
respectively, as listed in the table below. When a parameter is se lected the corresponding
indicator is highlighted. Turning [ADJ] knob will change the paramet er.

Catagory

Parameters

Ch 1
Vertical

Ø

Sensitivity

Ø

Position

Ø

Coupl e

Ch 2
Vertical

Ø

Sensitivity

Ø

Position

Ø

Couple

Selecting Button

[CH 1]

[S/DIV]

[CH 2]

Horizontal

Ø

Time base

Ø

Position

Trigger

Ø

Level

Ø

Sl ope

Ø

Source

Ø

Mode

[TRIGGER]

Adjustment

Turning
[ADJ]

Turning
[ADJ]

Turning
[ADJ]

Turning
[ADJ]

Remarks

Couple is directly set by slide switch at
the left side of panel and not controlled
by [CH 1] button.

3. Special Functions

Couple is directly set by slide switch at
the left side of panel and not controlled
by [CH 1] button.

Functions

Toggle HOLD/RUN state

Button Action

Pressing [HOLD]

Pressing [MENU]

Explantions

Freezing waveforms for static veiwing

Bring up Function Menu

Use menu to switch modes, save/recall waveforms, etc.

LCD backlight ON/OFF

Holding [ALT]

Fast adjustment ON/OFF

Pressing [ALT] at
specific cussor
locations

When vertical position (Ch 1 & Ch 2), horizontal position,
or trigger level is highlighted(selected) pressing [ALT] will
turn on Fast Adjustment, which increases the adjustment
incremental to 5 (or 25 for H. position) from default value 1

Persistent display ON/OFF

Pressing [ALT] when
cursor is at Timebase

Clear waveforms

Holding [S/DIV]

Clearing the whole waveform buffer as well as display

Ch 1 0V level alignment

Holding [CH 1]

Aligning 0V level of Ch 1 to its position indicator

Ch 2 0V level alignment

Holding [CH 2]

Aligning 0V level of Ch 2 to its position indicator

The table below lists some frequently used spec ial functions and related button actions.

4. About Rolling Display Mode

When Timebase is set to 0.1s/Div or slower DSO 094 enters

. In

this mode traces shift (rolling) from right to left across screen to display slow signals.

Rolling Display Mode

Connect battery

5. Calibration of 10X probe

Y channel as
trigge r source

IMPORTANT

Maximun allowed input voltages
are 50Vpk for 1X probe and 400Vpk
for 10X probe.

WARNING !

DO NOT ATTEMPT TO USE THE
DEVICE TO MEASURE LIVE
LINE VOLTAGE DIRECTLY!!!.

Rev. 01

Please refer to the technical note "

" (www.jyetech.com/

Support/HowToCalibrate10xProbe.pdf) for details.

How to Calibrate 10X Probe