Advanced Operations

Battery Running Time

The Built-in Battery Charger

Firmware Upgrade

DSO 094 uses one 3.7V 1200mAh Li-ion battery. When fully charged it can last
about 4 hours. If LCD backlight is turned off it can last longer.

The built-in charger is programmed to charge battery at current about 100mA. It
requires about 16 hours to fully charge a completely discharged battery. To charge
battery connect the USB port to PC or power adapter with USB connector. The
battery sign will blink indicating charging is under going.

From time to time firmware may need to be upgraded for new functions or
improved performance. This can be done by the built-in bootloader. For how the
bootloader works and how to use it please refer to

www.jyetech.com/Products/

LedScope/e094.php

Set up USB Connection

To use USB connection PC driver for CP2102 needs to be installed. The driver can be
downloaded at

.

www.silabs.com/products/mcu/pages/usbtouartbridgevcpdrivers.aspx.

Windows driver is also available at

www.jyetech.com/Support/Drivers&Tools.php

Save Captured Waveform

Press [MENU], scroll to "SAVE WAVEFORM" and press [HOLD]. Follow screen
instructions to save waveform to selected buffer.

Recall Saved Waveform

Send Screen as Bitmap File

1 ) Connect USB cable. Run a PC application that can handle XModem Protocol

(Windows HyperTerminal, for example). Set communication format to

and prepare it for file receiving.

38400bps,

8 data bits, 1 stop bit, no parity, no flow control

2 ) Adjust captured waveform so that the interested portion is displayed on screen.
3 ) Enter MENU and execute "SEND SCREEN".

You will see screen as shown in photo at right.

Restore Factory Default Settings

Press [MENU], scroll to "RECALL WAVEFORM" and press [HOLD]. Follow
screen instructions to recall saved waveform from selected buffer. Recalled wave-
form is displayed under HOLD state.

4 ) You can either start sending by first pressing

[HOLD] button then enabling PC receiving.
Or you can first enable PC receiving then press
[HOLD] to start sending. We recommend you
use the former sequence because it usually
results in shorter wait time.

Change Record Length (waveform buffer size)

Change Trigger Position (Pre-trig Length)

Change Trigger Sensitivity

Press [MENU], scroll to "RECALL WAVEFORM" and press [HOLD]. Follow
screen instructions to select record length desired.

Trig position is where the trig point is over the whole waveform buffer (see figure
below). You may like to set trig position closer to buffer start or to buffer end
depending on which portion of waveforms you are more interested.

In DSO 094 trig position is represented as percentage of buffer size. To change it
press [MENU], scroll to "CHANGE TRIG POS" and press [HOLD].

Trig sensitivity is the minimum level required to produce trigs. In DSO 094 it is
expressed as number of one tenth of major screen division size.To change it p

Default value is 4.

ress

[MENU], scroll to "CHANGE TRIG SEN" and press [HOLD]. The adjustable
range is 2 - 40.

Pre-trig Length

Buffer Size

Trig position

Press [MENU], scroll to "RESTORE DEFAULT" and press [HOLD]. Default values
are listed in the table at next page.

Maintenance

Hint: The reboot function under Menu is a quick way to enter the bootloader.

Powered From USB

When connectting the device to USB good quality cable is strongly recommended.
Bad cable could create too much voltage drop and consequently the device may
not be able to work properly.

Parameter

Range

Time base

H. Position

Ch 1 Sen.

Ch 1 Couple

Ch 1 V. Pos.

Ch 2 Sen.

Trigger Slope

Trigger Level

Trigger Src.

Trigger Pos.

Ch 2 Couple

Ch 2 V. Pos.

Trigger Mode

Trigger Sen.

Record Length

Working Mode

10M/div -- 0.2us/div

0 -- (Rec. Len. - 100)

2V/div -- 10mV/div

DC, AC, GND

-127 -- +127

Auto, Normal, Single

Falling, Rising

Ch 1, Ch 2

0% -- 100%

2 -- 40

500 -- 8000 points

Y/T, Y/X

Default

-127 -- +127

2V/div -- 10mV/div

DC, AC, GND

-127 -- +127

1ms/div

0

10mV/div

10mV/div

10

-10

Auto

Falling

0

Ch 1

50%

4

1000

Y/T

Parameter Ranges and Defaults

Channel Configuration

DSO 094 can be set to display two channels or one channel only by the Channel
Configuration Switch at top-left corner of panel. When it is set to one channel only
the not-selected channel is not displayed. However, its signal is still captured.

Number of Channel

2

Max. Sampling Rate

Analog Bandwidth

Max. Sensitivity

Resolution

Max. Record Length

Max. Sampling Rate

Fastest Time Base

Max. Input Voltage

Input Impedance

Num. of Waveform Mem.

Screen Size

Power Supply

Current Consumption

Dimension

Weight

50MSa/s

0 -- 10MHz

10mV/div

50Vpk (1X probe), 400Vpk(10X probe)

1M ohm/20pF

8 bits

8000 points

0.2us/div

4

128 X 64

Battery/USB

~300mA

140 x 70 x 30mm

190g

Trigger Modes

Trigger Position

Auto, Normal, Single

0% -- 100%

Specifications

Built-in Test Signal

The built-in test signal is a square wave signal with fixed frequency of 1KHz and
amplitude of about 1V peak-to-peak..

Technical Support:

Forum: http://forum.jyetech.com

Email: [email protected]