Guidance and manufacturer’s declaration – electromagnetic immunity
The model CL-R813 is intended for use in the electromagnetic immunity environment (Home
healthcare environment defined in IEC 60601-1-2) specified below.
The customer or the user of the model CL-R813 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
–guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines 100 kHz
±1 kV for input/output
lines 100 kHz
±2 kV for power
supply lines 100 kHz
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0 %
U
T
(100 % dip in
U
T
)
0.5 cycles and
1 cycle
70 %
U
T
(30 % dip in
U
T
)
25/30 cycles
0 %
U
T
(100 % dip in
U
T
)
250/300 cycles
0 %
U
T
(100 % dip in
U
T
)
0.5 cycles and
1 cycle
70 %
U
T
(30 % dip in
U
T
)
25/30 cycles
0 %
U
T
(100 % dip in
U
T
)
250/300 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
model CL-R813
requires continued
operation during
power mains
interruptions, it is
recommended that the
model CL-R813 be
powered from an
uninterruptible power
supply or a battery.
Power
frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
The product should be
kept at least 15 cm
away from the source
of power frequency
magnetic fields during
use.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание CL-R813
Страница 2: ... 2 All other products and company names are trademarks of their respective owners ENGLISH ...
Страница 13: ... 13 ENGLISH Label Important Please read the label containing warnings and cautions carefully before use ...
Страница 40: ... 40 ENGLISH External dimensionsor single screen 196 5 61 6 240 487 5 582 5 435 217 742 Unit mm ...
Страница 50: ... 2 日本語 記載されている会社名および商品名は各社の登録商標または商標です ...
Страница 78: ...ケー ブルマネジメント ケー ブルの接続時には スタンドのケー ブルホールを使用してまとめて配線してください ケーブルホール ケーブル 30 日本語 ...
Страница 82: ...外形図 単位 mm 196 5 61 6 240 487 5 582 5 435 217 742 34 日本語 ...