9 Technical data
86
9.3
Test voltages
9.4
Electrical safety
9.5
Analog output
Current
4 to 20 mA
0 to 20 mA
is detected
is detected
is detected
is not detected
is detected
is not detected
Input and output against voltage supply
- With voltage supply AC 110 to 240 V +10%, -15%
4260 V
- With voltage supply DC 24 V
4260 V
Clearances / creepage distances
Mains voltage to electronic com-
ponents and probes
≥
8 mm
/
≥
4 mm
Mains voltage to relays
≥
8 mm
/
≥
4 mm
Relays to electronic components
and probes
≥
8 mm
/
≥
4 mm
Resolution D/A converter >15 bit
Load resistance R
load
Accuracy
Burden influence
Voltage DC 0(2) to 10 V
Current DC 0(4) to 20 mA
≥
500
Ω
≤
500
Ω
≤ ±0.05% referring to 10 V
≤ ±0.05% referring to 20 mA
≤
±
15 mV
≤ ±0.02 %/100
Ω
Measuring probe
Out of range
Probe/cable break
Probe/cable short circuit