21
4 Mounting
4.4
Pressure connection
Seals
DANGER!
A flat seal, e.g. according to DIN EN 837, must be used for the appropriate process connection
for the ignition protection type intrinsic safety!
The pressure connection must be checked for seal tightness once established!
If the pressure transmitter is screwed into a threaded hole, the
full length
of the thread of the pres-
sure transmitter must be used!
NOTE!
The correct tightening torque is dependent on the size, material, and form of the seal used, as well as
the pressure connection of the pressure transmitter.
4.5
Cable connection inside terminal head
1. Press down the connection terminal (A) with a small screwdriver.
2. Insert the conductor into the opening, release connection terminal.
(A)
Connection terminal
(
A
)
Содержание dTRANS p33
Страница 2: ......
Страница 4: ...Weitere Informationen und Downloads qr 404753 de jumo info...
Страница 40: ...38 9 Baumusterpr fbescheinigung 9 Baumusterpr fbescheinigung...
Страница 41: ...39 9 Baumusterpr fbescheinigung...
Страница 42: ...9 Baumusterpr fbescheinigung 40...
Страница 43: ...41 9 Baumusterpr fbescheinigung...
Страница 45: ......
Страница 48: ...Further information and downloads qr 404753 en jumo info...
Страница 84: ...38 9 Examination certificate 9 Examination certificate...
Страница 85: ...39 9 Examination certificate...
Страница 86: ...9 Examination certificate 40...
Страница 87: ...41 9 Examination certificate...
Страница 89: ......
Страница 92: ...Informations compl mentaires et t l chargements qr 404753 fr jumo info...
Страница 128: ...38 9 Certificat d examen de type CE 9 Certificat d examen de type CE...
Страница 129: ...39 9 Certificat d examen de type CE...
Страница 130: ...9 Certificat d examen de type CE 40...
Страница 131: ...41 9 Certificat d examen de type CE...
Страница 133: ......
Страница 135: ......
Страница 136: ......