Julabo FD200 Скачать руководство пользователя страница 9

OPERATING PROCEDURES 

20 

 

Accessory: Clamp for cooler probe FT200/400 - order no. 8 970 400 

 

5.3. 

Tube connection FD200 

 

 

  Connect the tubes and secure with tube clamps. 

 

 

discharge (9)  

 

 

intake (10) 

Recommended flow rate: 

2 to 3 l/min 

 
 
 
 

6. 

OPERATING PROCEDURES 

6.1. 

Power connection 

 

Caution: 

  Only connect the unit to a power socket with earthing contact (PE – 

protective earth)! 

  The power supply plug serves as safe disconnecting device from the line 

and must be always easily accessible. 

  Never operate equipment with damaged mains power cables. 

  Regularly check the mains power cables for material defects (e.g. for 

cracks). 

  We disclaim all liability for damage caused by incorrect line voltages! 

 

 

Make sure that the line voltage and frequency match the supply 
voltage specified on the type plate. 
Deviations of ±10 % are permissible. 

 
 
 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание FD200

Страница 1: ...02 05 2016 English Operating manual Immersion Coolers FT200 FT400 FT900 Flow Through Cooler FD200 FT900 FT 400 1 951 4616 V4 05 16 www GlobalTestSupply com Find Quality Products Online at sales Globa...

Страница 2: ...5 V A 3 0 3 0 FT400 FT900 Temperature range C 40 30 90 30 Cooling capacity medium ethanol C kW 20 10 20 40 0 45 0 36 0 14 0 03 20 10 40 80 0 3 0 27 0 2 0 07 Cooling compressor 1 stage 2 stage Refriger...

Страница 3: ...fluorinated greenhouse gases Among other things it regulates the emission limits use and recovery of these substances It also contains requirements for operators of systems which require contain thes...

Страница 4: ...switch illuminated 2 Cooling control light 3 Clamp for immersion probe 7 Removable ventilation grid 8 Castor with locking lever 9 Tube connection discharge 10 Tube connection intake 11 Mains power cab...

Страница 5: ...OPERATING CONTROLS AND FUNCTIONAL ELEMENTS 16 FT900 FT900 1 2 3 7 8 Rear view 11 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 6: ...ly dangerous situation If these instructions are not followed serious injury and danger to life could result Caution Describes a possibly dangerous situation If this is not avoided slight or minor inj...

Страница 7: ...he air intake and exhaust grids free of obstructions Maintain a sufficient distance from all surrounding surfaces Do not move the unit from the position where it was set up during operation Always tur...

Страница 8: ...R404A 1 m 3 of space is required For 0 68 kg of refrigerant R23 1 m3 of space is required The ambient temperature must not exceed 35 C Keep at least 20 cm of open space on the front and rear venting...

Страница 9: ...tive earth The power supply plug serves as safe disconnecting device from the line and must be always easily accessible Never operate equipment with damaged mains power cables Regularly check the main...

Страница 10: ...ate The immersion probe as part of the cooling circuit should not be exposed to bath temperatures above the working temperature of the immersion cooler This would cause damage to the compressor Do not...

Страница 11: ...ays turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures Service and repair work may be performed only by authorized electrici...

Страница 12: ...returning a JULABO instrument for repair please contact an authorized JULABO service station When returning the unit Clean the unit in order to avoid any harm to the service personnel Attach a short f...

Отзывы: