End cycle: When sterilization is finished, cold process starts. The
lamp (6) is blinking and the lamp (4) lights on. The display (7) shows
the internal real temperature.
In case of solids sterilization fast steam output can be made using
the manual valve (16), the steam output is made with output (20)
YOU CAN NOT OPEN THE LID UNTIL THE PRESSURE IS NOT
0 BAR AND THE INTERNAL TEMPERATURE UNDER 95ºC. SEE
SECTION 8.1
When the lid is ready to be opened the display (7) shows the word
«End» and the lamp (6) is off.
If liquids or culture media are sterilized must leave cooler until the
pressure be 0 bar. During the cooling, is the same case of manual
depressurize, the display shows the internal temperature.
When the pressure is 0, the temperature is under 95ºC, the display
(7) shows the word «End». Therefore, the valve (20) can be opened
by the handle (16) and open the lid with his handle (17).
DO NOT OPEN THE VALVES 15 & 16 UNTIL THE
CYCLE IS COMPLETELY FINISHED TO AVOID
POSSIBLE BURNS WITH HEAT WATER OR
STEAM.
¡WARNING!
- THE DIRECT CONTACT WITH PRESSURE
STEAM CAN PRODUCE SERIOUS BURNS.
- IF YOU USE A HOSE TO MAKE THE STEAM
OUTPUT, DO NOT HOLD IT WITH THE HANDS
TO AVOID BURNS.
7