31

Table 4

Electromagnetic Compatibility Information

Recommended separation distances between portable and mobile RF communications equipment and 
the device

The device is intended for use in an electromagnetic environment in which radiated therefore 
disturbances are controlled. The customer or the user of the device can help prevent electromagnetic 
interference by maintaining a minimum distance between portable and mobile RF communications 
equipment (transmitters) and the device as recommended below, according to the maximum output 
power of the communications equipment.

Rated maximum output power 
of transmitter
           
           W

Separation distance according to frequency of transmitter

m

80 MHz to 800 MHz

800 MHz to 2.7 GHz
 

0.01

0.1

1

10

100

0.12

0.38

1.2

3.8

12

0.23

0.73

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in metres (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to 
the transmitter manufacturer.

NOTE1 At 80 MHz and 800 MHz, the separation distance for the higer frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.

2.3

7.3

23

1.  Probe

2.  START Button

3.  Set  Button

4.  Battery Cover 

Содержание DET-3011b

Страница 1: ...ended as a substitute for consultation with your physician 4 some parts are small enough to be swallowed 5 Never immerse this device in water or other liquids not waterproof 6 Do not modify this equip...

Страница 2: ...d with the memory function andhelp you to recognize each measurement result Please see the tinghe first use The thermometer has been designed for practical use It s not meant to replace a visit to the...

Страница 3: ...wait 10 minutes before taking a reading 4 Remove dirt from the forehead before taking a measurement Wait at least 10 minutes after washing the forehead area before taking a reading Temperature Taking...

Страница 4: ...iting period of 1 seconds between testing is recommended to ensure complete accuracy Note2 If the object to be tested is too far from the probe a prompt will appear See figure 5 To measure forehead te...

Страница 5: ...your thermometer will automatically be updated when it s connected with your phone 2 Confirm that your thermometer is connected successfully When your thermometer is connected successfully to your sma...

Страница 6: ...heric Pressure Clinical accuracy Dimension Weight Battery Battery life Approx 1 year 6000 readings Measuring site Forehead Forehead Mode Oral This thermometer converts the forehead temperature to disp...

Страница 7: ...mometer or to report unexpected operation or events The product is in compliance with the requirements of MDD 93 42 EEC 0197 is the identification number of notify body 0197 European Authorized Repres...

Страница 8: ...de 2 kV common mode N A Voltage dips short interrupti ons and voltage variations on p ower supply in put lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycle 70 UT 30 di...

Страница 9: ...ent Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 80 MHz to 800 MHz 800 MHz to 2 7 GHz 0 01 0 1 1 10 100 0 12 0 38 1 2 3 8 12 0 23 0 73 For tr...

Отзывы: