JOYTECH DET-1015b Скачать руководство пользователя страница 5

17

1.  The right ear reading may differ from the reading taken at the 

     left ear. Therefore, always take the temperature in the same ear.

2.  The ear must be free from obstructions or excess earwax buildup

     to take an accurate reading.

3.  External factors may influence ear temperatures such as:

       lying on one ear or the other

       their ears covered

       exposed to very hot or very cold temperatures

       recently swimming or bathing   

      In these cases, remove the individual from the situation and wait 

      20 minutes prior to taking a temperature.

4.  For persons wearing hearing aids or ear plugs, remove the device

     and wait 20 minutes prior to taking a temperature.

5.  When using the thermometer on infants under age 1, pull the 

     ear up making sure the sensor faces the eardrum.

     (See Figure 5)

6.  When using the thermometer on individuals over the age of 1, 

      pull the ear back making sure the sensor faces the eardrum. 

      (See Figure 6)

7.  Do not touch the probe window.  A cotton swab and rubbing 

     alcohol may be used to clean the sensor window.

8.  If the thermometer is stored in a significantly different 

     environment than testing location, place it in the testing 

     location for approximately 30 minutes prior to use.

15

16

Temperature Taking Hints

18

Temperature Taking Hints

Figure 5

Figure 6

    

Bluetooth requirements

The thermometer requires a bluetooth device with:

    . Bluetooth 4.0 or later

    . Android 6.0 or later

    . IOS 8.0 or later 

And works with:

     . iphone , iPod, iPad

     . Android Phones and Tablets

    Using for the first time

1. Download the "JoyHeath" App from Website or APP  Store (Such 

    as Apple Store).

2.Open the App on your phone or tablet. If requested, you should

     enable Bluetooth on your device. You can enable Bluetooth under 

     the Settings menu on your  smart phone or tablet.

3. Create a new user login, or login with your existing user name and 

    password.  

4. Selection device  “Thermometer”.

Illustration For Use

    

Match your thermometer with a Smart Device

1. If this is your first time using it, bind first.Open

 "

SETTING

"

 

    menu,choose

Bind and unbind device" and select the 

    appropriate model.

    The date and time on your thermometer will automatically be

     updated when it's connected with your phone.

2. Confirm that your thermometer is connected successfully.

    When your thermometer is connected successfully to your smart

     phone,  the “      ” symbol stop flashing and keep showing.

    

Transfer your readings

1. As soon as your measurement is finished, open the app on

     your smart phone to transfer  the reedings.

     Note: On the matched smart phone, Bluetooth must be enabled.

2. You can view your  temperature  readings in the app.

Illustration For Use

Illustration For Use

Product Description

1.  Probe Cover

2.  Probe

3.  Test Button

4.         Button

5.  Battery Cover

6.  Setting Button 

3

1

2

5

3

4

6

Electromagnetic Compatibility Information

Electromagnetic Compatibility Information

Содержание DET-1015b

Страница 1: ... doctor if you have health concerns 2 There is no gender or age limitation for using infrared ear thermometer 3 To avoid inaccurate readings always attach a new clean probe cover for each temperature reading 4 Do not expose the thermometer to temperature extremes below 25 13 or over 55 131 nor excessive humidity 95 RH 5 This thermometer must only be used with this brand thermometer probe covers 6 ...

Страница 2: ...clock setting section to learn how to setup the time in the first use The thermometer has been designed for practical use It s not meant to replace a visit to the doctor Please also remember to compare the measurement result to your regular body temperature Please see the section to learn how to measure the body temperature Illustration For Use Beep Alarm Memory Mode There are each 10 sets memorie...

Страница 3: ...he thermometer Set the unit Press Button to select the unit you want After the unit is set press Setting Button the time format figure will appear Set the time format The device can display the time in either an AM PM 12 hour or a 24 00 24 hour format Press and release button to select the format With the preferred time format on the display press Setting Button the Hour figure is flashing automat...

Страница 4: ... the probe firmly making sure the probe and the inner portion of the probe cover fit snugly 2 Press the The display is activated to show all segments Remarks a Probe cover sign stand for taking the ear temperature with probe cover now and vice versa b To avoid the risk of cross contamination it is recommended that you always measure the ear temperature with probe cover To achieve accurate readings...

Страница 5: ...re Taking Hints 18 Temperature Taking Hints Figure 5 Figure 6 Bluetooth requirements The thermometer requires a bluetooth device with Bluetooth 4 0 or later Android 6 0 or later IOS 8 0 or later And works with iphone iPod iPad Android Phones and Tablets Using for the first time 1 Download the JoyHeath App from Website or APP Store Such as Apple Store 2 Open the App on your phone or tablet If reque...

Страница 6: ...ollowed by Figure 2 4 or finished testing press and hold the Button for three seconds The letter MEM will appear at the bottom of the display See Figure 5 2 The thermometer will automatically memorize the last 30 temperature readings Each time the Button is pressed the screen displays past readings that correspond with a number 1 30 The number 1 reflects the most recent reading while the number 30...

Страница 7: ...9 6 In Object mode Temperature taken is lower than 0 32 23 Storage and Transportation Temperature Limit 13 131 25 55 25 55 Specifications Ear Object mode 10 40 50 104 15 85 RH non condensing 00hPa 1060hPa Atmospheric Pressure 7 57 146 38mm Approx 83grams with batteries DC3V 2 AA battery Measuring range Laboratory accuracy Display resolution 0 1 or 0 1 Measure time Approximately one second Operatin...

Страница 8: ...dards And the manufacturer is ISO 13485 certified Thermometer is warranted by manufacture to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery to the first user who purchases the instrument This warranty does not cover batteries damage to the probe window or damage to the instrument caused by misuse negligence or accide...

Страница 9: ... conditions described in the table below The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communications equipment can affect the device Use of the unit in conjunction with non approved accessories can affect the device negatively and alter the electromagnetic c...

Страница 10: ...4 MHz 7 MHz to 7 3 MHz 10 1 MHz to 10 15 MHz 14 MHz to 14 2 MHz 18 07 MHz to 18 17 MHz 21 0MHz to 21 4MHz 24 89 MHz to 24 99 MHz 28 0 MHz to 29 7 MHz and 50 0 MHz to 54 0 MHZ b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 7 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause inte...