Unit Operation 

Unit Operation 

Unit Operation 

Low Battery Indicator

The unit will broadcast "Low Battery" when battery life is depleting and unable to inflate cuff for 
testing.  The "             " appears simultaneously for approximately 5 seconds prior to shutting 
off.  Replace batteries at this time.  No memory loss will occur throughout this process.

Unit Operation 

19

In the power down state, press and hold the " START/STOP" button,  and then install the batteries.  Until the LCD 

screen is full, release the " START/STOP" button. When the LCD screen displays the  double zero, the blood 

pressure meter is in static state. Software version is displayed :10 is a software version in the figure.

Static Pressure Measurement

Note: Only Service personnel permitted to access to this mode, the  mode unavailable in 
            normal use.

Bluetooth requirements

   The monitor requires a device with:

   . Bluetooth 4.0 or later

   . Android 5.0 or later

   . IOS 9.0 or later 

   And works with:

   . iphone , iPod, iPad

   . Android Phones and Tablets

-Pairing

 your monitor with a 

S

mart Device

            Wireless communication

             Frequency range : 2.4 Ghz (2400-2483.5 MHz)

             Modulation : GFSK

             Antenna gain:0.5dBi

 

1. Download the free “JoyHealth”

 

App: On your mobile phone or table 

go to www.sejoy.com.

2.Open the App on your phone or tablet. If requested, you should enable Bluetooth on your

 device. You can enable Bluetooth under the Settings menu on your smart phone or table.

3. Create a new user login, or login with your existing user name and password.  

4. Selection device  “Blood pressure monitor”.

Bluetooth connection

-Using

 

for

 

the

 

first

 

time

-Transfer your readings

1. As soon as your measurement is complete, open the app on your smart phone to transfer

   your reedings.

Notr: On the paired smartphone, Bluetooth must be enabled.

2. You can view your blood pressure readings on the app.

   

Note:
1 )If another family member uses it, clean the cuff before use,and useit after it dries naturally; 
2) Press “SET" to set the memory group, set the memory group.
3 )Restart the measurement according to the operation method.
4) If you want to view the measured value, please refer to the memory view description on page 18.

1. Long press the "memory" button (or the "M" key) on your monitor until the word "boo" flashes 

on your monitor .

The MAC address

Bluetooth

SMART

AM

2. Open the APP and click "pair device" button, a popup window will appear. Click "Search device"
 button to display all sphygmomanometer devices(The display information includes the device model 
and MAC address), then you can find the sphygmomanometer through the "device Name", click the 
pictureA film or text can be used to pair the device.

3. After the pairing is complete, the device interface switches from boo interface, indicating that
 the pairing is successful.

Troubleshooting

Abnormal 

phenomenon

Processing method

Abnormal 

sphygmomanometer

Cause analysis

Note: If you cannot solve the abnormal situation by yourself, you can consult the manufacturer or 
the manufacturer's designated unit by phone. It is forbidden to disassemble and repair without 
permission. If necessary, professional maintenance personnel can ask the manufacturer for the list 
of components and circuit schematic diagram.

The following table shows the error signs that may occur during measurement, possible causes 
and handling methods. Please measure again using the correct method

Er1

Er2

Er3

Er4

Error display

The cause of the problem

The solution

Can't detect high and low 

pressure

Cuff too loose or loose

Severe arm or body movement

 leads to abnormal reporting errors

The pressure exceeds 300mmHg

Please fasten the cuff before 

measuring

Hold the arm or body still and

 measure again

Replace the battery or connect the

 power adapter (if any).

Er5

The pressure exceeds 15mmHg 

for 3 minutes

Check whether the cuff is knotted or the 

vent valve is blocked. If the problem

 persists, contact the manufacturer

Move the arm during measurement or

Electronic sphygmomanometer

Speaking, nervous or emotional

 during measurement

There is interference in charging process 

or improper operation in measuring 

process

 

Incorrect measurement posture

Stay quiet, keep your arm steady, and 
do not move the monitor

Instead of talking, take deep
 breaths to calm your mood and 
relax your body

Adjust posture, see 
"Blood pressure gauge Wearing”

See operation Instructions.

The armband is tied too tight or too loose,

Or the arm strap is tied incorrectly;

Roll the armband correctly

Please fasten the cuff before 

measuring

Please fasten the cuff before 

measuring

Содержание DBP-6293L

Страница 1: ...C 81060 2 30 for electronic sphygmomanometers Precautions to Ensure Safe Reliable Operation 1 Do not drop the unit Protect it from sudden jars or shocks 2 Do not insert foreign objects into any openin...

Страница 2: ...f the performance of this equipment could result 34 the blood pressure monitors equipped with bluetooth BT module is only transmit historical blood pressure BP from the subject device to a user suppli...

Страница 3: ...etplace comfortably back and arm support on a desk or table withyour legs uncrossed your arm resting on a firm and your feet flat on the floor See Figure C 5 Apply cuff to your left arm and middle of...

Страница 4: ...format ON or OFF by pressing the MEM button 8 Saved Settings 7 Volume Setting Press SET button to enter volume setting mode Set the voice volume by adjusting the MEM button There are six volume levels...

Страница 5: ...re in arm cuff becomes too extreme while testing press the START STOP button to turn power off The cuff pressure will rapidly dissipate once the unit is off Testing 1 Power On Press and hold START STO...

Страница 6: ...mory button or the M key on your monitor until the word boo flashes on your monitor The MAC address Bluetooth SMART AM 2 Open the APP and click pair device button a popup window will appear Click Sear...

Страница 7: ...course of a day Many factors including diet stress cuff placement etc may affect an individual s blood pressure Q Should I apply the cuff to the left or right arm What is the difference A Either arm c...

Страница 8: ...fications are subject to change without notice Bandwidth 2 0 MHz Safety Standard included but not limited 1 IEC 80601 2 30 medical electrical equipment part 2 30 particular requirements for the basic...

Страница 9: ...r frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50Hz or 60Hz Power frequency magnetic fields should be at levels charactertic of a typical location in a typical comme rcial or hospital enviro...

Страница 10: ...product leaves the factory Patients can use the method described in the section Verify Manometer Pressure Accuracy to verify the accuracy If the accuracy deviation is large please contact the manufact...

Отзывы: