Unit Operation
Unit Operation
Unit Operation
Low Battery Indicator
The unit will broadcast "Low Battery" when battery life is depleting and unable to inflate cuff for
testing. The " " appears simultaneously for approximately 5 seconds prior to shutting
off. Replace batteries at this time. No memory loss will occur throughout this process.
Unit Operation
19
In the power down state, press and hold the " START/STOP" button, and then install the batteries. Until the LCD
screen is full, release the " START/STOP" button. When the LCD screen displays the double zero, the blood
pressure meter is in static state. Software version is displayed :10 is a software version in the figure.
Static Pressure Measurement
Note: Only Service personnel permitted to access to this mode, the mode unavailable in
normal use.
Bluetooth requirements
The monitor requires a device with:
. Bluetooth 4.0 or later
. Android 5.0 or later
. IOS 9.0 or later
And works with:
. iphone , iPod, iPad
. Android Phones and Tablets
-Pairing
your monitor with a
S
mart Device
Wireless communication
Frequency range : 2.4 Ghz (2400-2483.5 MHz)
Modulation : GFSK
Antenna gain:0.5dBi
1. Download the free “JoyHealth”
App: On your mobile phone or table
go to www.sejoy.com.
2.Open the App on your phone or tablet. If requested, you should enable Bluetooth on your
device. You can enable Bluetooth under the Settings menu on your smart phone or table.
3. Create a new user login, or login with your existing user name and password.
4. Selection device “Blood pressure monitor”.
Bluetooth connection
-Using
for
the
first
time
-Transfer your readings
1. As soon as your measurement is complete, open the app on your smart phone to transfer
your reedings.
Notr: On the paired smartphone, Bluetooth must be enabled.
2. You can view your blood pressure readings on the app.
Note:
1 )If another family member uses it, clean the cuff before use,and useit after it dries naturally;
2) Press “SET" to set the memory group, set the memory group.
3 )Restart the measurement according to the operation method.
4) If you want to view the measured value, please refer to the memory view description on page 18.
1. Long press the "memory" button (or the "M" key) on your monitor until the word "boo" flashes
on your monitor .
The MAC address
Bluetooth
SMART
AM
2. Open the APP and click "pair device" button, a popup window will appear. Click "Search device"
button to display all sphygmomanometer devices(The display information includes the device model
and MAC address), then you can find the sphygmomanometer through the "device Name", click the
pictureA film or text can be used to pair the device.
3. After the pairing is complete, the device interface switches from boo interface, indicating that
the pairing is successful.
Troubleshooting
Abnormal
phenomenon
Processing method
Abnormal
sphygmomanometer
Cause analysis
Note: If you cannot solve the abnormal situation by yourself, you can consult the manufacturer or
the manufacturer's designated unit by phone. It is forbidden to disassemble and repair without
permission. If necessary, professional maintenance personnel can ask the manufacturer for the list
of components and circuit schematic diagram.
The following table shows the error signs that may occur during measurement, possible causes
and handling methods. Please measure again using the correct method
Er1
Er2
Er3
Er4
Error display
The cause of the problem
The solution
Can't detect high and low
pressure
Cuff too loose or loose
Severe arm or body movement
leads to abnormal reporting errors
The pressure exceeds 300mmHg
Please fasten the cuff before
measuring
Hold the arm or body still and
measure again
Replace the battery or connect the
power adapter (if any).
Er5
The pressure exceeds 15mmHg
for 3 minutes
Check whether the cuff is knotted or the
vent valve is blocked. If the problem
persists, contact the manufacturer
Move the arm during measurement or
Electronic sphygmomanometer
Speaking, nervous or emotional
during measurement
There is interference in charging process
or improper operation in measuring
process
Incorrect measurement posture
Stay quiet, keep your arm steady, and
do not move the monitor
Instead of talking, take deep
breaths to calm your mood and
relax your body
Adjust posture, see
"Blood pressure gauge Wearing”
See operation Instructions.
The armband is tied too tight or too loose,
Or the arm strap is tied incorrectly;
Roll the armband correctly
Please fasten the cuff before
measuring
Please fasten the cuff before
measuring