Unit Operation 

Unit Operation 

Unit Operation 

Unit Operation 

11

System Settings

With power off, press "SET" button to activate System Settings.  The Memory Group icon 

flashes.

1.  Select Memory Group 

     While in the System Setting mode, you may accumulate test  results into 2 different 

     groups.  This allows multiple users to save individual test results (up to 150 memories 

     per group.)  Press " MEM "  button to choose a group setting.  Test results will 

     automatically store in each selected group.

2.  Time/Date setting
     Press

SET

 button again to set the Time/Date mode. Set the year first by adjusting 

     the 

MEM

 button.

     Press

SET

 button again to confirm current month. Continue setting the dat, hour 

     and minute in the same way. Every time the 

SET

 button is pressed, it will lock 

     in your selection and continue in succession ( month,day,hour, minute,12/24 hours)

AM

AM

3. Time Format setting
     Press 

 SET 

button again to set the time fornat setting mode.Set the time format 

     by adjusting the

MEM

 button. EU means European Time US means U.S Time

6.Voice Setting
Press "SET" button to enter voice setting mode. Set voice format ON or OFF by
pressing the "MEM" button.

8.  Saved Settings

7.Volume Setting
Press "SET" button to enter volume setting mode. Set the voice volume by adjusting
the "MEM" button . There are six volume levels.

     While in any setting mode, press " ON/OFF " button to turn the unit off.  All 
     information will be saved.  

Note:  If unit is left on and not in use for 3 minutes, it will automatically save all information 

            and shut off.

Applying the Arm Cuff

1.  Pick up the sphygmomanometer, open the arm band, put the sphygmomanometer and arm 

band on the arm, to firm air leakage is appropriate;

2.  With sticky nylon section facing outward, insert end of cuff underneath metal ring 

     of cuff  

3.  Fasten cuff about 

above the elbow joint.  For best results apply cuff to 

     bare arm and keep level with heart while testing

.

1-2cm (0.4-0.8'') 

.

Note:  Do not insert air plug into opening located on right side of monitor unit.  

            This opening is designed for an optional power supply only.

1-

2c

''

(0

.4

-0

.8

4.Triple measurement function
Press "SET" to enter the triple measurement function setting mode, and press "MEM" to 
turn ON or OFF the triple measurement function.

5.

 time

Press "SET" to enter the setting mode of three measurement intervals. By pressing the 
"MEM" to switch three measurement intervals, respectively "30", "60" and "90".

Triple measurement

Содержание DBP-6293L

Страница 1: ...C 81060 2 30 for electronic sphygmomanometers Precautions to Ensure Safe Reliable Operation 1 Do not drop the unit Protect it from sudden jars or shocks 2 Do not insert foreign objects into any openin...

Страница 2: ...f the performance of this equipment could result 34 the blood pressure monitors equipped with bluetooth BT module is only transmit historical blood pressure BP from the subject device to a user suppli...

Страница 3: ...etplace comfortably back and arm support on a desk or table withyour legs uncrossed your arm resting on a firm and your feet flat on the floor See Figure C 5 Apply cuff to your left arm and middle of...

Страница 4: ...format ON or OFF by pressing the MEM button 8 Saved Settings 7 Volume Setting Press SET button to enter volume setting mode Set the voice volume by adjusting the MEM button There are six volume levels...

Страница 5: ...re in arm cuff becomes too extreme while testing press the START STOP button to turn power off The cuff pressure will rapidly dissipate once the unit is off Testing 1 Power On Press and hold START STO...

Страница 6: ...mory button or the M key on your monitor until the word boo flashes on your monitor The MAC address Bluetooth SMART AM 2 Open the APP and click pair device button a popup window will appear Click Sear...

Страница 7: ...course of a day Many factors including diet stress cuff placement etc may affect an individual s blood pressure Q Should I apply the cuff to the left or right arm What is the difference A Either arm c...

Страница 8: ...fications are subject to change without notice Bandwidth 2 0 MHz Safety Standard included but not limited 1 IEC 80601 2 30 medical electrical equipment part 2 30 particular requirements for the basic...

Страница 9: ...r frequency 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50Hz or 60Hz Power frequency magnetic fields should be at levels charactertic of a typical location in a typical comme rcial or hospital enviro...

Страница 10: ...product leaves the factory Patients can use the method described in the section Verify Manometer Pressure Accuracy to verify the accuracy If the accuracy deviation is large please contact the manufact...

Отзывы: