![Joycare JC-615 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/joycare/jc-615/jc-615_manual_2027554056.webp)
56
57
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT-
GUIDANCE
conducted RF
IEC 61000-4-6
3Vrms
150kHz to
80MHz
3Vrms
Portable and mobile communication
equipment should be used no closer to any
part of the model JC-615(NB-152U) including
cables, than the recommended separation di-
stance calculated from the equation applicable
to the frequency of the transmitter.
Radiated RF
IEC 61000-4-3
10V/m
80MHz
to2.5GHz
3V/m
Recommended separation distance
Where P is the maximum output power
rating of transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in meters
(m). Field strengths fixed RF transmitter
as determined by an electromagnetic site
survey* should be less than the compliance
level in each frequency range.
*Interference may occur in the vicinity of
equipment marked with following symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation is effected
by absorption and reflection from structures, objects and people.
* field strength from fixed transmitters such as base stations for radio (cellar/cordless) telepho-
nes ,land mobile radio, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To access to electromagnetic environment duo to fixed RF
transmitter, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the model JC-615(NB-152U) is used exceeds the applicable RF complian-
ce level above, the model JC-615(NB-152U) should be observed, to very normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Model
* Over the frequency range 150 kHz to 80MHz field strengths should be less than 3 V/m