MANUFACTURER’S GUIDE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The clinical thermometer JC-132G (FDTH-V0-7) can be used in the following electromagnetic
environment. The customer or user of the clinical thermometer JC-132G (FDTH-V0-7) must
make sure that the device is used in said environment.
EMISSION TEST
COMPLIANCE
GUIDE TO THE ELECTROMAGNETIC
ENVIRONMENT
Electrostatic discharges (ESD)
IEC/EN61000-4-2
± 6kV in contact
± 8kV in the air
The floors should be made of
wood, cement or baked clay. If
the floors are covered in synthe-
tic material, relative humidity
should not exceed 30%.
Magnetic field
IEC/EN 61000-4-8
3 A/m
The magnetic field should
be the typical magnetic field
of a commercial or hospital
environment.
Irradiated immunity
IEC/EN61000-4-6
3 V/m 80 MHz at 2.5 GHz
(in appliances that are not
life equipment)
Содержание JC-132G