background image

-22- 

 

Jovy profiles -> Export profile to RE-7550

 

 

Export a saved profile in PC to be exported to machine memory 
and Run the profile without using PC from the machine 
interface direct.  
 

 

Jovy profiles -> Import profile From RE-7550

 

 

Import a profile saved in Machine memory to be saved on PC 
to be run from PC Suite. 
 

Note: PC Suite Run only profiles saved to PC 
 

 

Run Manual control Process

 

 

 

Select Run Manual Control process from main Interface 
window  

 

 

Adjust the Heaters power level to start the process 

 

Set the Alarm warning temperature value (when the TC read 
that value a beep will start to work) if no need for this option 
ignore click the box.  

 

Содержание RE-7550

Страница 1: ...JOVY SYSTEMS RE 7550 User Manual Rev 1 00...

Страница 2: ...mer 3 Specifications 4 Safety Caution instructions 4 RE 7550 hardware description 6 PC Suite Version 2 00 installation 6 Operation through Machine interface 7 Guide to create Process profile 13 Jovy P...

Страница 3: ...and Liability Disclaimer The RE 7550 package including any hard copies and soft copies are copyright protected The purchase of the programs confers the right to make back up copies of the software in...

Страница 4: ...s voltages applied to them Failing to follow the operating instructions could result in death serious injuries and or extensive damage to property Never open the housing or one of the Infrared heaters...

Страница 5: ...surface When the Infrared heater is in use its housing and the areas exposed to the infrared rays can reach high level of temperature Please do not touch the housing until it cooled down Any maintenan...

Страница 6: ...er Pre Heater 3 Thermocouple 4 LCD 5 X Y Table 6 Cooling Fan 7 Laser Pointer 8 Guard Glass 9 Control Switches 10 Power Plug 11 Power Switch 12 Fuses 13 Thermocouple Plug 14 USB Port 15 Picking Up Tube...

Страница 7: ...s Used to change the value Up to increase and Down to decrease as well to set the Lower Heater operating power level percentage from 0 to 100 OK Menu used to open menu and also to select any item insi...

Страница 8: ...computer Run Jovy PC Suite the machine automatically run on USB mode Cooling fan directed toward the application Laser pointer points the target inside the Heating area Target located in the middle o...

Страница 9: ...aters switch off and external fan switch on Temperature measurement unit Celsius or Fahrenheit Activate or de activate the security password to lock the machine access Change the default password 9999...

Страница 10: ...h reflective tape to prevent Infrared rays to reach it Pre heat until the PCB reaches a temperature from 80 C to 120 C Adjust the upper heater to the desired power level then rotate the upper heater s...

Страница 11: ...s available to save up to 50 profiles into Machine memory Option Three Delete a saved profile from machine memory Option Four Delete all saved profiles from machine memory Create a Process Profile Fro...

Страница 12: ...age Heaters Power level Step 5 Yes to save the profile into Machine memory No to Run the profile process Step 6 save the profile to its location according to the machine profiles counter Step 7 Name t...

Страница 13: ...icated algorithm to end each phase successfully this is implemented in the three phases preheat soak and re flow Use the Thermo couple in bended end method as shown below The success time for a profil...

Страница 14: ...to 3 UH power value is 80 LH power value is 100 and preheats end temperature value is 120 C Soaking UH level set to 3 UH power value is 90 LH power value is 75 and preheats end temperature value is 2...

Страница 15: ...CD drive then windows OS folder o Run dotnetfx exe file then follow the install instructions appeared on your screen Install Microsoft Visual C Runtime Redistributable o Run CD drive then windows OS...

Страница 16: ...n in Figures below Jovy PC Suite on Offline mode no any connection Detected Jovy PC Suite detected RE 7550 and displays the machine serial number NOTE Jovy PC Suite software should be run as an admini...

Страница 17: ...Main Interface Temperature value as measured from Thermocouple Machine status connected Online Dis connected Offline Thermocouple status Connected Dis connected Machine mode Park Normal mode Time scal...

Страница 18: ...ide lines Tools Machine settings Temperature measurement Unit C F Interface Language English Chinese Safety Set point immediate Switch off of heaters and operates the cooling Fan Product ID RE 8500 or...

Страница 19: ...well time on Peak temperature TAL 4 Set the warning alarm temperature 5 Write a message to be displayed while the stage is running 6 Set the Heaters power levels for each profile stage 7 Save the crea...

Страница 20: ...le Editor Edit Profile 1 Import any profile saved on PC to Edit one or more parameter 2 Import any profile saved on Machine memory up to 50 profiles After editing Three options are available same as 7...

Страница 21: ...s Jovy Profile Manager Delete a profile saved to machine memory by its number Format and delete all profiles saved to machine memory Jovy Profile Manager is a tool to Edit re arrange or delete one or...

Страница 22: ...7550 Import a profile saved in Machine memory to be saved on PC to be run from PC Suite Note PC Suite Run only profiles saved to PC Run Manual control Process Select Run Manual Control process from ma...

Страница 23: ...erature and process time unit and then process will start once Ok pressed While Manual control process is running heaters can be stopped or heaters power level can be changed then apply changes for th...

Страница 24: ...24 Run a Profile Process Run Profile only from saved profile to PC Choose the selected profile then Run it Import a profile saved to machine memory and Run it with Run Profile Process...

Страница 25: ...Warranty policy The complete machine has a warranty period of 1 year from the time of purchase We provide online troubleshooting support and technical advice service through the web site www jovy sys...