36
D Technical Specifications
Physical
Size
198mm (L)×197mm (W)×67mm (H)
Weight
(include
battery and
pads)
≤2kg
Environmental
Temperature
and relative
humidity
Operating(with battery and pads installed)
0
℃~
50
℃
,5%
~
95% RH(non-condensing)
Standby (with battery and pads installed)
5
℃~
35
℃
,5%
~
95% RH(non-condensing)
Storage/shipping(with battery and pads )
-30
℃~
65
℃
for up to 24 hours, 5%
~
95%
RH(non-condensing)
Altitude
0 to 15000 feet
Sealing
IP54
Vibration
Operating: meets EN1789 random, road ambulance.
Standby: meets EN1789 swept sine, road ambulance.
shock/drop
Withstands 1 meter drop to any edge, corner, or surface.
EMC
Refer to Annex G for EMC Information
Defibrillator
Waveform
Biphasic truncated exponential. Waveform parameters are
automatically adjusted to patients
'
impedances.
Energy
150 J nominal (±15%) delivered into a 50 ohm load.
When the system detects impedance ≤ 20Ω or ≥ 200Ω,
device will not deliver a shock.
Charge time
from
shock-advised
≤7s with a new battery
Pads
Pads type
Disposable, pre-gelled self-adhesive defibrillation pads, latex
free
Cable Length
120cm±12cm
Active gel
surface area
100cm
2
±10cm
2
each
Pads shelf life
2 years from date of manufacture when stored and
maintained according to directions provided in this manual.
Battery
Battery type
12VDC,4.2 Ah, LiMnO
2
battery
Battery
capacity
New battery≥200 shocks (20
℃
±2
℃
).
New battery≥100 shocks (0
℃
±2
℃
).
Содержание iAED-S1