4.3.5 ECG Rhythm Analysis
Procedure
Keep the patient
’s body position steady and wait for the ECG Rhythm analysis
completion.
Warning:
Do not touch or shake the patient during the rhythm analysis
otherwise it will affect the rhythm analysis results.
Details
※
The defibrillator analyzes the patient
’s ECG signal and determines whether
the rhythm is shockable or nonshockable. Meanwhile the AED will continue to
monitor the pads connections and will abort analysis if it detects any
connection problems.
⚫
If the iAED-S1 has determined that the pads are
poorly contacted to the patient, the device will
prompts “Poor pad contact to patient, press
pads firmly, check connector” The user should
place pads correctly on the patient following
instructions on pad package.
⚫
If t
he pads are properly placed, the defibrillator will prompt “Do not touch
patient! Analyzing heart rhythm.” At this time, the user should not touch
the patient.
⚫
If the iAED-S1 has detected motion in the patient the defibrillator will stop
analysis for up to 10 seconds. The defibrillator notifies the responder of
the problem. The defibrillator will resume analysis after 10 seconds, even
if motion is still present.
⚫
Rhythm analysis takes about 10 seconds. Observe the patient through the
whole process. If the patient is awake, stop the defibrillation immediately
by pressing the ON/OFF button to turn off the defibrillator.
Warning:
Do not touch or shake the patient during the rhythm analyzing