background image

3

 

 

 

Getting  Started 

 

 

Remove  the unit from the  box and lay it on a 

at surface. 

 

1)

 

Connect the female end of the power cord to the rear of the unit, and the 
male end in a grounded electrical outlet.

 

2) 

The  male  jack  of  the  Black (or Green)  surgery hand  piece is inserted into 
the unit into the black socket marked 

“ HANDPIECE”

  on the front  panel.

 

3) 

I

ndiffe

rent  Plate:  This is the red cord.  The male jack should be inserted 

into  the red female  jack opening  marked

 

“IND PLATE” on the front 

panel. 

4) 

Connect the foot pedal in the rear of the unit.

 

5) 

The Indifferent Plate should be placed under the animal close to 
the procedure site. Conductive gel must be used to create uniform 
skin contact and maximize surface contact with the plate. 

Failure 

to use gel can result in plate burns to the animal 

in addition to 

poor or no function. 

 

6) 

Select an electrode and insert the straight shaft end into the top 
opening of the handpiece. Be careful not to bend or damage the 
electrode in any way. Do not attempt to unscrew the handpiece 
as this will damage it. 

 

7) 

Set the waveform dial to the desired mode.

 

8) 

Select the power intensity. 1 is lowest and 10 is highest.  Always 
start at the minimum level and gradually work up to a level where 
cutting is smooth but no sparks are emitted.

 

9) 

The “ACTIVE” light should be 

lit when the footswitch is pressed. This 

indicates that radio frequency waves are being sent.

 

10)    The  electrode  should  be  gently  held  on  the  tissue.  A  smooth, 

constant  motion  will  produce  less  lateral  heat  and  the  cleanest 
incision.  The  tissue  should  always  be  kept  moist  for  better  and 
cleaner cutting. Cutting should be done without pressure, as this 
can damage the tissue and tip.  

Содержание ELECTROSURGE J0470M

Страница 1: ...J0470M ELECTROSURGE Radio Frequency Electrosurgery Unit Operation Instructions ...

Страница 2: ...ELECTROSURGE J0470M Hand Piece Connection Black Socket Hand Piece Indifferent Plate Connection Red Socket Indifferent Plate Intensity Level Waveform Mode Electrodes Foot Switch Power Active Light ...

Страница 3: ...The high frequency radio waves pass through tissue and make a precise surgical incision just like a scalpel blade The surgeon chooses from varying radio waveforms that result in varying degrees from pure cutting to hemostasis Radio frequency electrosurgery should not be confused with electrocautery Electrocautery uses thermal energy or heat that simply burns the tissue The advantages of electrosur...

Страница 4: ...e preferred mode for thin avian skin and incisions where heavy bleeding is nota problem This mode provides for the least amount of hemostasis Coag 1 mode This selection produces a periodic sine wave with brief rests between cycles This waveform has less cutting power but coagulation of small capillary beds is achieved Coag 2 mode This selection produces a partially rectified sine wave This provides...

Страница 5: ...e gel can result in plate burns to the animal in addition to poor or no function 6 Select an electrode and insert the straight shaft end into the top opening of the handpiece Be careful not to bend or damage the electrode in any way Do not attempt to unscrew the handpiece as this will damage it 7 Set the waveform dial to the desired mode 8 Select the power intensity 1 is lowest and 10 is highest A...

Страница 6: ...top of the indifferent plate 3 Select an electrode and place it in the hand piece 4 Turn waveform mode to Coag 2 5 Set power level at 8 position 6 Press on the footpedal 7 Do several incisions and observe the results This high setting will cause sparking and may char tissue 8 Adjust power to 1 This will illustrate poor cutting or extreme drag 9 Move the power setting to different levels and continu...

Страница 7: ...not covered with blood The waveform should bein Coag 1 or Coag 2 Coag 2 is preferred for larger bleeders The two types of coagulation electrodes are the round ball electrodes and heavier needle An indirect method uses a hemostat with a fine tip The vessel is grasped by the hemostat The electrode is applied to the hemostat about 1 from its end Care and Selection of Electrodes The electrodes can beau...

Страница 8: ...JorVet Electrode is the industry standard 3 32 shank Other manufacturer s electrodes will work on the JorVet unit GT1 Blade Electrode GT2 Small Loop Electrode GT3 Large Loop Electrode GT4 Fine Wire Electrode GT5 Heavy Wire Electrode GT8 Small Ball 2 38mm Electrode GT9 Heavy Ball 5mm Electrode GT1 GT2 GT3 GT4 GT5 GT8 GT9 ...

Страница 9: ...nsult a physician before exposing anyone wearing a pacemaker to radio frequency waves 2 Various ECG and monitoring equipment can be affected by radio frequency waves Place these items as far away from the Indifferent Plate as possible 3 When changing electrodes do not have the foot pedal depressed 4 Electrosurgery can give off sparks and ignite flammable liquids or gases 5 Electrosurgery still requir...

Страница 10: ...tcut a Is the power light on b Foot pedal connected c ACTIVE light on when foot pedal is depressed d Electrode properly attached e Indifferent Plate attached and properly placed Autoclaving Handpiece Electrodes The handpiece and electrodes should be thoroughly cleaned and dried following a procedure These can then be autoclaved in a sterilization pouch at no more than 250 F for 10 15 minutes ...

Страница 11: ...9 SKU J0470M Specifications 110 volt available in 220v Output power 100 watts Output frequency 2 0 mHz Size Height 3 Width 8 Depth 9 Weight 7 lbs Replacement Fuse 12 5a 250v fuse ...

Страница 12: ...______________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ Users Notes ...

Страница 13: ...against defects in workmanship and materials The completed Warranty Card must be returned to Jorgensen Laboratories Inc within 14 days of purchase Handpiece 1 year Indifference Plate 90 days Foot Switch 90 days Electrodes are not covered under warranty Pocket for Copy of Purchase Invoice ...

Страница 14: ...Distributed By Jorgensen Laboratories 1450 North Van Buren Avenue Loveland CO 80538 800 525 5614 FAX 970 663 5042 www jorvet com info jorvet com ...

Отзывы: