Wandering Baseline
Appears on the ECG as a fluctuation of the tracing up and
downward on the grid.
Causes:
• Dirty or corroded electrodes
• Loose electrodes or electrodes positioned on a bony area
• Insufficient or dried out Alcohol or Electrode Gel
• Rising and falling of chest during normal or apprehensive
respiration
When wandering baseline occurs:
• Clean skin with alcohol or acetone if necessary
• Reposition electrodes
• Check the electrode connections
• Assist the patient in relaxation
Other
When an electrosurgical unit is used with the cardiograph, noise
generated by an ESU is superimposed on the ECG waveforms.
Equipment in the same room as the cardiograph may generate
noise which is superimposed on the ECG waveforms.
• Noise generated by an electrosurgical unit
• Noise from AC power line
• Spike noise caused by electrostatic energy as shown below
.